Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes
- PMID: 16501412
- DOI: 10.1097/01.opx.0000200673.96758.7b
Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes
Abstract
Purpose: The ICare (Tiolat Oy, Helsinki, Finland) is a new portable tonometer that measures intraocular pressure (IOP) with a new rebound method, in which a very light probe is used to make momentary contact with the cornea in slow motion. The purpose of this study is to assess the accuracy of the ICare IOP measurements by comparing them against other portable tonometers: Perkins applanation tonometer and Tono-Pen XL digital tonometer (Medtronic Solan, Jacksonville, FL).
Methods: Sixty-five young subjects were assessed with each of the tonometers. ICare tonometry was performed first, followed by Perkins applanation tonometry and Tono-Pen XL in a random order. Regression analysis was used to evaluate the relationship between the Perkins tonometer and the remaining tonometers used in this study. Tonometers were also compared by plotting the difference between the methods against the mean. The hypothesis of zero bias was examined by a paired t test and the 95% limits of agreement (LoA) were calculated.
Results: ICare and Tono-Pen XL significantly overestimate IOP when compared with Perkins applanation tonometry. The mean of the difference between Perkins and ICare and Perkins and Tono-Pen XL was (mean +/- standard deviation) -3.35 +/- 2.28 mm Hg and -2.78 +/- 2.53 mm Hg, respectively. The 95% LoA between Perkins tonometry and ICare tonometry were between -7.81 and +1.12 and between Perkins tonometry and Tono-Pen XL tonometry between -7.74 and +2.18.
Conclusions: Compared with Perkins tonometry, the ICare tonometer allows clinicians to estimate IOP with a portable, rapid, and noninvasive method with similar reliability to that offered by Tono-Pen XL. Clinicians should be aware of the systematic overestimation of IOP with the ICare. Further research is needed to evaluate the performance of rebound tonometry in populations with higher IOP and assess the reliability of this technique in the early detection and follow up of glaucomatous patients.
Similar articles
-
[Comparative study of two portable tonometers: Tono-Pen XL and Perkins].Arch Soc Esp Oftalmol. 2003 Apr;78(4):189-96. Arch Soc Esp Oftalmol. 2003. PMID: 12743842 Spanish.
-
Comparison of intraocular pressure measurements between the Tono-Pen XL® and Perkins® applanation tonometers in dogs and cats.Vet Ophthalmol. 2012 Mar;15 Suppl 1:14-20. doi: 10.1111/j.1463-5224.2011.00926.x. Epub 2011 Jul 12. Vet Ophthalmol. 2012. PMID: 22050624
-
Comparison of the rebound tonometer (ICare) to the applanation tonometer (Tonopen XL) in normotensive dogs.Vet Ophthalmol. 2006 Jan-Feb;9(1):17-21. doi: 10.1111/j.1463-5224.2005.00429.x. Vet Ophthalmol. 2006. PMID: 16409240 Clinical Trial.
-
Meta-analysis of the concordance of Icare® PRO-based rebound and Goldmann applanation tonometry in glaucoma patients.Eur J Ophthalmol. 2020 Mar;30(2):245-252. doi: 10.1177/1120672119866067. Epub 2019 Aug 29. Eur J Ophthalmol. 2020. PMID: 31466475
-
Update in tonometry. Phosphene and rebound tonometries, self-tonometry and technologies for the future.Bull Soc Belge Ophtalmol. 2007;(303):87-95. Bull Soc Belge Ophtalmol. 2007. PMID: 17894294 Review.
Cited by
-
Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.Vet Ophthalmol. 2021 Mar;24 Suppl 1(Suppl 1):175-185. doi: 10.1111/vop.12832. Epub 2020 Oct 17. Vet Ophthalmol. 2021. PMID: 33070466 Free PMC article.
-
Intraocular Pressure Measurement in Childhood Glaucoma under Standardized General Anaesthesia: The Prospective EyeBIS Study.J Clin Med. 2022 May 18;11(10):2846. doi: 10.3390/jcm11102846. J Clin Med. 2022. PMID: 35628971 Free PMC article.
-
Comprasion of ICare rebound tonometer and Goldmann applanation tonometer in high myopia.ScientificWorldJournal. 2014;2014:869460. doi: 10.1155/2014/869460. Epub 2014 Oct 14. ScientificWorldJournal. 2014. PMID: 25379549 Free PMC article. Clinical Trial.
-
Age differences in central and peripheral intraocular pressure using a rebound tonometer.Br J Ophthalmol. 2006 Dec;90(12):1495-500. doi: 10.1136/bjo.2006.103044. Epub 2006 Aug 2. Br J Ophthalmol. 2006. PMID: 16885185 Free PMC article.
-
Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation.J Ophthalmol. 2019 May 9;2019:1964107. doi: 10.1155/2019/1964107. eCollection 2019. J Ophthalmol. 2019. PMID: 31210982 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources