Influence of disease severity and optic disc size on the diagnostic performance of imaging instruments in glaucoma
- PMID: 16505035
- DOI: 10.1167/iovs.05-1133
Influence of disease severity and optic disc size on the diagnostic performance of imaging instruments in glaucoma
Abstract
Purpose: To evaluate the influence of disease severity and optic disc size on the diagnostic accuracy of three imaging technologies for structural assessment in glaucoma: confocal scanning laser ophthalmoscopy, scanning laser polarimetry, and optical coherence tomography.
Methods: One hundred five patients with glaucoma and 61 normal subjects were recruited from the Diagnostic Innovations in Glaucoma Study (DIGS). All individuals underwent imaging with the GDx Variable Corneal and Lens Compensator (VCC; Carl Zeiss Meditec, Inc., Dublin, CA), the Heidelberg Retina Tomograph II (HRT II; Heidelberg Engineering, GmbH, Dossenheim, Germany), and the Stratus OCT (Stratus OCT; Carl-Zeiss Meditec, Inc.) within a 6-month period. Severity of disease was based on the AGIS (Advanced Glaucoma Intervention Study) visual field score. To evaluate the influence of severity of glaucoma and optic disc size on the diagnostic accuracy of the imaging instruments, the sensitivities of the tests were fitted as a function of the AGIS score and disc area, by using logistic marginal regression models.
Results: The severity of visual field loss had a significant influence on the sensitivity of all imaging instruments. More severe disease was associated with increased sensitivity. This influence was similar among the three instruments. With regard to optic disc area, larger optic discs were associated with decreased sensitivity for the Stratus OCT parameter Average Thickness and the GDx VCC parameter Nerve Fiber Indicator, whereas small optic discs were associated with increased sensitivity. For the HRT II parameter Moorfields regression analysis classification, an inverse effect was observed.
Conclusion: The diagnostic performances of the GDx VCC, HRT II, and Stratus OCT were significantly influenced by the severity of the disease and optic disc size. These covariates should be taken into account when comparing the performances of these tests for glaucoma diagnosis.
Similar articles
-
Effect of optic disc size and disease severity on the diagnostic capability of glaucoma imaging technologies in an Indian population.J Glaucoma. 2012 Sep;21(7):475-80. doi: 10.1097/IJG.0b013e31821829f1. J Glaucoma. 2012. PMID: 21522023
-
Glaucoma diagnostics.Acta Ophthalmol. 2013 Feb;91 Thesis 1:1-32. doi: 10.1111/aos.12072. Acta Ophthalmol. 2013. PMID: 23384049
-
Effect of glaucomatous damage on repeatability of confocal scanning laser ophthalmoscope, scanning laser polarimetry, and optical coherence tomography.Invest Ophthalmol Vis Sci. 2007 Mar;48(3):1156-63. doi: 10.1167/iovs.06-0921. Invest Ophthalmol Vis Sci. 2007. PMID: 17325159 Free PMC article.
-
Retinal nerve fiber layer analysis in the diagnosis of glaucoma.Curr Opin Ophthalmol. 2006 Apr;17(2):120-31. doi: 10.1097/01.icu.0000193079.55240.18. Curr Opin Ophthalmol. 2006. PMID: 16552246 Review.
-
Image analysis of optic nerve disease.Eye (Lond). 2004 Nov;18(11):1207-13. doi: 10.1038/sj.eye.6701544. Eye (Lond). 2004. PMID: 15534606 Review.
Cited by
-
Optic nerve head parameters of high-definition optical coherence tomography and Heidelberg retina tomogram in perimetric and preperimetric glaucoma.Indian J Ophthalmol. 2016 Apr;64(4):277-84. doi: 10.4103/0301-4738.182938. Indian J Ophthalmol. 2016. PMID: 27221679 Free PMC article.
-
The papilla as screening parameter for early diagnosis of glaucoma.Dtsch Arztebl Int. 2008 Aug;105(34-35):583-9. doi: 10.3238/arztebl.2008.0583. Epub 2008 Aug 25. Dtsch Arztebl Int. 2008. PMID: 19471619 Free PMC article.
-
The location of the inferior and superior temporal blood vessels and interindividual variability of the retinal nerve fiber layer thickness.J Glaucoma. 2010 Mar;19(3):158-66. doi: 10.1097/IJG.0b013e3181af31ec. J Glaucoma. 2010. PMID: 19661824 Free PMC article.
-
Understanding disparities among diagnostic technologies in glaucoma.Arch Ophthalmol. 2012 Jul;130(7):833-40. doi: 10.1001/archophthalmol.2012.786. Arch Ophthalmol. 2012. PMID: 22776920 Free PMC article.
-
Retinal nerve fibre thickness measured with optical coherence tomography accurately detects confirmed glaucomatous damage.Br J Ophthalmol. 2007 Jul;91(7):905-7. doi: 10.1136/bjo.2006.111252. Epub 2007 Feb 14. Br J Ophthalmol. 2007. PMID: 17301118 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical