Optimized contrast-enhanced CT protocols for diagnostic whole-body 18F-FDG PET/CT: technical aspects of single-phase versus multiphase CT imaging
- PMID: 16513616
Optimized contrast-enhanced CT protocols for diagnostic whole-body 18F-FDG PET/CT: technical aspects of single-phase versus multiphase CT imaging
Abstract
The purpose of this study was to compare various PET/CT examination protocols that use contrast-enhanced single-phase or contrast-enhanced multiphase CT scans under different breathing conditions.
Methods: Sixty patients with different malignant tumors were randomized into 4 different PET/CT protocols. Single-phase protocols included an intravenous contrast-enhanced (Ultravist 370; iodine at 370 mg/mL) single-phase whole-body CT scan (90 mL at 1.8 mL/min; delay, 90 s) during shallow breathing (protocol A) or during normal expiration (NormExp; protocol B). Multiphase protocols included 2 separate CT scans in the arterial contrast enhancement phase (90 mL at 2.5-2.8 mL/min; bolus tracking; scan range, base of the skull to the kidneys) and the portal-venous contrast enhancement phase (delay, 90 s; scan range, base of the lungs to the proximal thighs) during shallow breathing (protocol C) or during NormExp (protocol D) followed by a low-dose CT scan during shallow breathing for attenuation correction and whole-body PET. Feasibility was assessed by comparing the misalignment of the upper abdominal organs quantitatively by means of the craniocaudal, lateral, and anterior-posterior differences on coregistered PET/CT images. For image quality, the occurrence of CT artifacts and mismatching of rigid body points were evaluated qualitatively.
Results: Misalignment was significantly lower for protocol B in almost all organs and represented the best coregistration quality. Surprisingly, protocol A showed significantly better alignment than the multiphase CT scans during NormExp. Misalignment values between the multiphase protocols were not significantly different, with a trend toward lower values for protocol D. The best CT image quality, with a significantly lower occurrence of artifacts, was found for protocols B and D (NormExp). The levels of mismatching of rigid body points because of patient movement in between the transmission and emission scans were similar for all protocols.
Conclusion: Multiphase CT protocols presented a technical disadvantage represented by suboptimal image coregistration compared with single-phase protocols. Nevertheless, multiphase protocols are technically feasible and should be considered for patients who will benefit from a contrast-enhanced multiphase CT examination for diagnosis.
Comment in
-
Adequate evaluation of image registration in hybrid PET/CT.J Nucl Med. 2006 Sep;47(9):1556; author reply 1557-8. J Nucl Med. 2006. PMID: 16954568 No abstract available.
Similar articles
-
Adequate evaluation of image registration in hybrid PET/CT.J Nucl Med. 2006 Sep;47(9):1556; author reply 1557-8. J Nucl Med. 2006. PMID: 16954568 No abstract available.
-
Comparison of whole-body PET/CT, dedicated high-resolution head and neck PET/CT, and contrast-enhanced CT in preoperative staging of clinically M0 squamous cell carcinoma of the head and neck.J Nucl Med. 2009 Aug;50(8):1205-13. doi: 10.2967/jnumed.109.062075. Epub 2009 Jul 17. J Nucl Med. 2009. PMID: 19617339
-
Two-dimensional vs three-dimensional imaging in whole body oncologic PET/CT: a Discovery-STE phantom and patient study.Q J Nucl Med Mol Imaging. 2007 Sep;51(3):214-23. Epub 2007 Apr 30. Q J Nucl Med Mol Imaging. 2007. PMID: 17464266
-
Whole-body magnetic resonance imaging and positron emission tomography-computed tomography in oncology.Top Magn Reson Imaging. 2007 Jun;18(3):193-202. doi: 10.1097/RMR.0b013e318093e6bo. Top Magn Reson Imaging. 2007. PMID: 17762383 Review.
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
Cited by
-
Putting 'clear' into nuclear medicine: a decade of PET/CT development.Eur J Nucl Med Mol Imaging. 2006 Aug;33(8):857-61. doi: 10.1007/s00259-006-0137-z. Eur J Nucl Med Mol Imaging. 2006. PMID: 16794821 No abstract available.
-
Multiphase contrast-enhanced CT with highly concentrated contrast agent can be used for PET attenuation correction in integrated PET/CT imaging.Eur J Nucl Med Mol Imaging. 2012 Feb;39(2):316-25. doi: 10.1007/s00259-011-1919-5. Epub 2011 Nov 29. Eur J Nucl Med Mol Imaging. 2012. PMID: 22124778
-
PET-CT in oncology: making the most of CT.Cancer Imaging. 2008 Oct 4;8 Spec No A(Spec Iss A):S87-93. doi: 10.1102/1470-7330.2008.9015. Cancer Imaging. 2008. PMID: 18852084 Free PMC article. Review.
-
Intravenous contrast-enhanced CT can be used for CT-based attenuation correction in clinical (111)In-octreotide SPECT/CT.EJNMMI Phys. 2015 Dec;2(1):3. doi: 10.1186/s40658-015-0108-1. Epub 2015 Feb 12. EJNMMI Phys. 2015. PMID: 26501805 Free PMC article.
-
Low dose non-enhanced CT versus standard dose contrast-enhanced CT in combined PET/CT protocols for staging and therapy planning in non-small cell lung cancer.Eur J Nucl Med Mol Imaging. 2007 Jan;34(1):36-44. doi: 10.1007/s00259-006-0186-3. Epub 2006 Aug 1. Eur J Nucl Med Mol Imaging. 2007. PMID: 16896664
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical