. 2006 Nov;144(1):296-307.
doi: 10.1016/j.envpol.2005.11.046. Epub 2006 Mar 3.

Calibration and field performance of membrane-enclosed sorptive coating for integrative passive sampling of persistent organic pollutants in water

Affiliations

Calibration and field performance of membrane-enclosed sorptive coating for integrative passive sampling of persistent organic pollutants in water

Branislav Vrana et al. Environ Pollut. 2006 Nov.

Abstract

Membrane-enclosed sorptive coating (MESCO) is a miniaturised monitoring device that enables integrative passive sampling of persistent, hydrophobic organic pollutants in water. The system combines the passive sampling with solventless preconcentration of organic pollutants from water and subsequent desorption of analytes on-line into a chromatographic system. Exchange kinetics of chemicals between water and MESCO was studied at different flow rates of water, in order to characterize the effect of variable environmental conditions on the sampler performance, and to identify a method for in situ correction of the laboratory-derived calibration data. It was found that the desorption of chemicals from MESCO into water is isotropic to the absorption of the analytes onto the sampler under the same exposure conditions. This allows for the in situ calibration of the uptake of pollutants using elimination kinetics of performance reference compounds and more accurate estimates of target analyte concentrations. A field study was conducted to test the sampler performance alongside spot sampling. A good agreement of contaminant patterns and water concentrations was obtained by the two sampling techniques.