. 2006 Mar 14;113(10):1305-10.
doi: 10.1161/CIRCULATIONAHA.105.602490. Epub 2006 Mar 6.

Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates

Affiliations

Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates

Jörg Hausleiter et al. Circulation. .

Abstract

Background: Multislice computed tomography angiography (CTA) is a promising technology for imaging patients with suspected coronary artery disease. Compared with 16-slice CTA, the improved spatial and temporal resolution of 64-slice CTA (0.6- versus 1.0-mm slice thickness and 330- versus 420-ms gantry rotation time) is associated with an increase in radiation dose. The objective of this retrospective investigation was to compare the estimated dose received during 16- and 64-slice CTA in daily practice and to investigate the impact of different scan protocols on dose and image quality.

Methods and results: Radiation dose was estimated for 1035 patients undergoing coronary CTA. Scanning algorithms with and without an ECG-dependent dose modulation and with a reduced tube voltage were investigated on dose estimates and image quality. In the entire patient cohort, radiation dose estimates were 6.4+/-1.9 and 11.0+/-4.1 mSv for 16- and 64-slice CTA, respectively (P<0.01). The reduction in radiation dose estimates ranged between 37% and 40% and between 53% and 64% with the use of ECG-dependent dose modulation and with the combined use of the dose modulation and a reduced tube voltage, respectively. The reduction in dose estimates was not associated with a reduction in diagnostic image quality as assessed by the signal-to-noise ratio and by the frequency of coronary segments with diagnostic image quality.

Conclusions: The increase in spatial and temporal resolution with 64-slice CTA is associated with an increased radiation dose for coronary CTA. Dose-saving algorithms are very effective in reducing radiation exposure and should be used whenever possible.