Non-invasive glucose monitoring in patients with diabetes: a novel system based on impedance spectroscopy
- PMID: 16524714
- DOI: 10.1016/j.bios.2006.01.031
Non-invasive glucose monitoring in patients with diabetes: a novel system based on impedance spectroscopy
Abstract
The aim of this work was to evaluate the performance of a novel non-invasive continuous glucose-monitoring system based on impedance spectroscopy (IS) in patients with diabetes. Ten patients with type 1 diabetes (mean+/-S.D., age 28+/-8 years, BMI 24.2+/-3.2 kg/m(2) and HbA(1C) 7.3+/-1.6%) and five with type 2 diabetes (age 61+/-8 years, BMI 27.5+/-3.2 kg/m(2) and HbA(1C) 8.3+/-1.8%) took part in this study, which comprised a glucose clamp experiment followed by a 7-day outpatient evaluation. The measurements obtained by the NI-CGMD and the reference blood glucose-measuring techniques were evaluated using retrospective data evaluation procedures. Under less controlled outpatient conditions a correlation coefficient of r=0.640 and a standard error of prediction (SEP) of 45 mg dl(-1) with a total of 590 paired glucose measurements was found (versus r=0.926 and a SEP of 26 mg dl(-1) under controlled conditions). Clark error grid analyses (EGA) showed 56% of all values in zone A, 37% in B and 7% in C-E. In conclusion, these results indicate that IS in the used technical setting allows retrospective, continuous and truly non-invasive glucose monitoring under defined conditions for patients with diabetes. Technical advances and developments are needed to expand on this concept to bring the results from the outpatient study closer to those in the experimental section of the study. Further studies will not only help to evaluate the performance and limitations of using such a technique for non non-invasive glucose monitoring but also help to verify technical extensions towards a IS-based concept that offers improved performance under real life operating conditions.
Similar articles
-
Non-invasive glucose monitoring in patients with Type 1 diabetes: a Multisensor system combining sensors for dielectric and optical characterisation of skin.Biosens Bioelectron. 2009 May 15;24(9):2778-84. doi: 10.1016/j.bios.2009.02.001. Epub 2009 Feb 20. Biosens Bioelectron. 2009. PMID: 19286364
-
In vivo glucose monitoring: the clinical reality and the promise.Biosens Bioelectron. 2005 Apr 15;20(10):1897-902. doi: 10.1016/j.bios.2004.08.016. Biosens Bioelectron. 2005. PMID: 15741056 Review.
-
Novel planar glucose biosensors for continuous monitoring use.Biosens Bioelectron. 2005 Apr 15;20(10):1993-2000. doi: 10.1016/j.bios.2004.09.010. Biosens Bioelectron. 2005. PMID: 15741068
-
Novel micromachined silicon sensor for continuous glucose monitoring.Biosens Bioelectron. 2005 Nov 15;21(5):802-8. doi: 10.1016/j.bios.2005.02.008. Biosens Bioelectron. 2005. PMID: 16242621
-
Home blood glucose biosensors: a commercial perspective.Biosens Bioelectron. 2005 Jun 15;20(12):2435-53. doi: 10.1016/j.bios.2004.11.012. Epub 2005 Jan 18. Biosens Bioelectron. 2005. PMID: 15854818 Review.
Cited by
-
Passive Diffusion of Transdermal Glucose: Noninvasive Glucose Sensing Using a Fluorescent Glucose Binding Protein.J Diabetes Sci Technol. 2014 Mar;8(2):291-298. doi: 10.1177/1932296813519994. Epub 2014 Jan 21. J Diabetes Sci Technol. 2014. PMID: 24876581 Free PMC article.
-
Retrospective Continuous-Time Blood Glucose Estimation in Free Living Conditions with a Non-Invasive Multisensor Device.Sensors (Basel). 2019 Aug 24;19(17):3677. doi: 10.3390/s19173677. Sensors (Basel). 2019. PMID: 31450547 Free PMC article.
-
The compensation of perturbing temperature fluctuation in glucose monitoring technologies based on impedance spectroscopy.Med Biol Eng Comput. 2007 Sep;45(9):863-76. doi: 10.1007/s11517-007-0229-3. Epub 2007 Jul 28. Med Biol Eng Comput. 2007. PMID: 17661101
-
Evaluation of a Continuous Blood Glucose Monitor: A Novel and Non-Invasive Wearable Using Bioimpedance Technology.J Diabetes Sci Technol. 2023 Mar;17(2):336-344. doi: 10.1177/19322968211054110. Epub 2021 Oct 29. J Diabetes Sci Technol. 2023. PMID: 34711074 Free PMC article.
-
Developing and evaluating non-invasive healthcare technologies for a group of female participants from a socioeconomically disadvantaged area.Sci Rep. 2021 Dec 13;11(1):23896. doi: 10.1038/s41598-021-03262-3. Sci Rep. 2021. PMID: 34903797 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous