Validation of TM-2655 oscillometric device for blood pressure measurement
- PMID: 16534410
- DOI: 10.1097/01.mbp.0000200484.49540.12
Validation of TM-2655 oscillometric device for blood pressure measurement
Abstract
Objective: To perform validation for an arm-type oscillometric TM-2655 device (A&D Company Ltd, Tokyo, Japan) for blood pressure measurement according to the British Hypertension Society protocol.
Methods: Eighty-five study participants (33 men and 52 women) were included in the study. Mean age was 52.9+/-15.0 years, systolic blood pressure range was 84-208 mmHg and diastolic blood pressure range was 48-120 mmHg. For each participant, three readings of TM-2655 were compared with sequential auscultatory measurements by two trained independent observers. The observers used a calibrated mercury sphygmomanometer and dual stethoscope. The results were graded according to the British Hypertension Society protocol 1993.
Results: The average difference between mercury sphygmomanometer and TM-2655 readings for systolic blood pressure was -1.0+5.2 mmHg (mean+/-SD) and for diastolic blood pressure -0.9+/-4.7 mmHg. The proportions of values agreeing to within 5, 10 and 15 mmHg were 72.5, 93.7 and 99.6% for systolic blood pressure and 78.8, 96.9 and 100% for diastolic blood pressure between the observers and the device (A/A British Hypertension Society grade).
Conclusions: The TM-2655 device achieved British Hypertension Society grade A/A and therefore can be recommended for blood pressure measurement in an adult population.
Similar articles
-
Validation of three oscillometric blood pressure devices against auscultatory mercury sphygmomanometer in children.Blood Press Monit. 2006 Oct;11(5):281-91. doi: 10.1097/01.mbp.0000209082.09623.b4. Blood Press Monit. 2006. PMID: 16932037 Clinical Trial.
-
Validation of the OMRON M7 (HEM-780-E) blood pressure measuring device in a population requiring large cuff use according to the International Protocol of the European Society of Hypertension.Blood Press Monit. 2007 Jun;12(3):173-8. doi: 10.1097/MBP.0b013e3280b08367. Blood Press Monit. 2007. PMID: 17496467
-
Simultaneous validation of the SunTech 247 diagnostic station blood pressure measurement device according to the British Hypertension Society protocol, the International Protocol and the Association for the Advancement of Medical Instrumentation standards.Blood Press Monit. 2009 Oct;14(5):222-7. doi: 10.1097/MBP.0b013e328330c873. Blood Press Monit. 2009. PMID: 19741510
-
Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysis.Blood Press Monit. 2019 Dec;24(6):265-276. doi: 10.1097/MBP.0000000000000412. Blood Press Monit. 2019. PMID: 31658107
-
Demise of the mercury sphygmomanometer and the dawning of a new era in blood pressure measurement.Blood Press Monit. 2003 Feb;8(1):19-21. doi: 10.1097/01.mbp.0000057012.67622.de. Blood Press Monit. 2003. PMID: 12604931 Review.
Cited by
-
Efficacy of a multi-component exercise programme and nutritional supplementation on musculoskeletal health in men treated with androgen deprivation therapy for prostate cancer (IMPACT): study protocol of a randomised controlled trial.Trials. 2017 Oct 3;18(1):451. doi: 10.1186/s13063-017-2185-z. Trials. 2017. PMID: 28974267 Free PMC article. Clinical Trial.
-
Mixed-methods feasibility study of blood pressure self-screening for hypertension detection.BMJ Open. 2019 May 29;9(5):e027986. doi: 10.1136/bmjopen-2018-027986. BMJ Open. 2019. PMID: 31147366 Free PMC article.
-
Association Between Neutrophil-to-Lymphocyte Ratio and Benign Prostatic Hyperplasia: Results from the TCLSIH Cohort Study.J Inflamm Res. 2023 Oct 25;16:4857-4866. doi: 10.2147/JIR.S431049. eCollection 2023. J Inflamm Res. 2023. PMID: 37904787 Free PMC article.
-
North Korean refugee health in South Korea (NORNS) study: study design and methods.BMC Public Health. 2012 Mar 8;12:172. doi: 10.1186/1471-2458-12-172. BMC Public Health. 2012. PMID: 22401814 Free PMC article.
-
The relationship between sodium concentrations in spot urine and blood pressure increases: a prospective study of Japanese general population: the Circulatory Risk in Communities Study (CIRCS).BMC Cardiovasc Disord. 2016 Mar 5;16:55. doi: 10.1186/s12872-016-0219-1. BMC Cardiovasc Disord. 2016. PMID: 26944259 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources