In vitro model for intraoperative adjustments in an implantable hearing aid (MET)
- PMID: 16540912
- DOI: 10.1097/01.MLG.0000199889.61483.8F
In vitro model for intraoperative adjustments in an implantable hearing aid (MET)
Abstract
Objective: Assessment of the optimal static preloading of Otologics Middle Ear Transducer (MET) Ossicular Stimulator, when coupled to the incus.
Background: The MET Ossicular Stimulator is a partially implantable electromagnetic middle ear hearing device that transmits vibrations to the ossicular chain. The vibration patterns were measured with laser-Doppler vibrometry.
Study design: Experimental.
Material: We used three human cadaveric temporal bones (TB) and one MET ossicular stimulator.
Methods: Laser-Doppler vibrometry was used for the selection of TBs. The cochlea was subsequently extirpated from the posterior side to measure the vibrational patterns (VP) of the footplate. Three TBs with different VP were selected based on data obtained from volunteers with normal hearing (n = 110): one TB with a VP larger than +1 SD, one TB with a VP in the range of +/-1 SD, and 1 TB with a VP smaller than -1 SD. Transfer functions were calculated between VP of the measurement points at the coupling rod, umbo, incus, and footplate. The TBs were subsequently defrosted. The MET was implanted and coupled to the ossicular chain. Different coupling loads were measured at the incus, the umbo, and the footplate.
Results: Optimal transfer function between the MET transducer and the oval window was achieved during contact when the coupling rod advanced 0.0625 mm (90 degrees rotation). Additional advances of 0.0625 mm (180 degrees turn = 0.125 mm) resulted in a decreased vibrational amplitude, ranging between 20 and 40 dB below 3 kHz. The lowest linear distortion occurred up to 10 kHz during direct contact without advancing the coupling rod.
Similar articles
-
Intraoperative ossicular loading with the Otologics fully implantable hearing device.Acta Otolaryngol. 2007 Apr;127(4):360-4. doi: 10.1080/00016480601089424. Acta Otolaryngol. 2007. PMID: 17453454
-
A novel implantable hearing system with direct acoustic cochlear stimulation.Audiol Neurootol. 2008;13(4):247-56. doi: 10.1159/000115434. Epub 2008 Feb 7. Audiol Neurootol. 2008. PMID: 18259077 Clinical Trial.
-
Anatomical vibrations that implantable microphones must overcome.Otol Neurotol. 2007 Aug;28(5):579-88. doi: 10.1097/01.mao.0000271700.52982.d8. Otol Neurotol. 2007. PMID: 17534199
-
Implantable hearing device performance measured by laser Doppler interferometry.Ear Nose Throat J. 1997 May;76(5):297-9, 302, 305-9. Ear Nose Throat J. 1997. PMID: 9170711 Review.
-
Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.Laryngoscope. 2006 Aug;116(8):1321-33. doi: 10.1097/01.mlg.0000230479.39551.4a. Laryngoscope. 2006. PMID: 16885731 Review.
Cited by
-
[Semi-implantable hearing aids for sensorineural hearing loss and combined hearing loss: experiences at the German Armed Forces Hospital in Ulm].HNO. 2009 Mar;57(3):208-15. doi: 10.1007/s00106-008-1851-6. HNO. 2009. PMID: 19247625 Review. German.
-
In situ Probe Microphone Measurement for Testing the Direct Acoustical Cochlear Stimulator.Front Neurosci. 2017 Aug 15;11:450. doi: 10.3389/fnins.2017.00450. eCollection 2017. Front Neurosci. 2017. PMID: 28860963 Free PMC article.
-
[Differential indication of active middle ear implants].HNO. 2015 Jun;63(6):402-18. doi: 10.1007/s00106-015-0018-5. HNO. 2015. PMID: 26054730 Review. German.
-
Implantable hearing devices.GMS Curr Top Otorhinolaryngol Head Neck Surg. 2017 Dec 18;16:Doc06. doi: 10.3205/cto000145. eCollection 2017. GMS Curr Top Otorhinolaryngol Head Neck Surg. 2017. PMID: 29279724 Free PMC article. Review.
-
Optimum Coupling of an Active Middle Ear Actuator: Effect of Loading Forces on Actuator Output and Conductive Losses.Otol Neurotol. 2019 Jul;40(6):789-796. doi: 10.1097/MAO.0000000000002253. Otol Neurotol. 2019. PMID: 30994569 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous