Initial experience with the enclose proximal aortic anastomosis device during off-pump coronary artery bypass: an alternative to aortic side clamping
- PMID: 16543162
Initial experience with the enclose proximal aortic anastomosis device during off-pump coronary artery bypass: an alternative to aortic side clamping
Abstract
Background: The utilization of off-pump coronary artery bypass surgery (OPCAB) has resulted in the development of new technology to facilitate the creation of aorto-coronary graft anastomoses. Proximal aortic devices (PADs) enable the construction of a proximal aortic anastomosis without the use of a side-biting aortic clamp, thus reducing the risk of neurologic injury from particulate embolization.
Methods: One hundred ninety-seven patients underwent OPCAB at our institution between January 2003 and December 2004. Sixty (30.5%) patients had proximal aorto-coronary graft anastomoses constructed with the Novare Enclose PAD. The remaining 137 (69.5%) patients had graft construction with a standard aortic side-clamp technique. We compared the outcomes of these 2 cohorts to evaluate the safety and efficacy of the Novare Enclose PAD.
Results: One hundred seven proximal anastomoses were constructed in the PAD group, and 199 proximal were constructed in the side-clamp group. Three patients (1.5%), all in the side-clamp group, sustained permanent neurologic deficits after OPCAB. There were 2 cases of device malfunctions. There were no anastomotic thromboses, no reoperations for anastomotic hemorrhage, and no patients required anastomotic revision. Of the 197 patients in the series, there were 4 deaths, 2 in each group, resulting in an overall mortality rate of 2%.
Conclusion: The Novare Enclose PAD is a safe device that facilitates suture construction of proximal aorto-coronary graft anastomosis. In a select group of patients, the use of this device may reduce the risk of neurologic injury when compared to the application of an aortic side-biting clamp for coronary bypass surgery.
Similar articles
-
Cerebrovascular Events After No-Touch Off-Pump Coronary Artery Bypass Grafting, Conventional Side-Clamp Off-Pump Coronary Artery Bypass, and Proximal Anastomotic Devices: A Meta-Analysis.J Am Heart Assoc. 2016 Feb 18;5(2):e002802. doi: 10.1161/JAHA.115.002802. J Am Heart Assoc. 2016. PMID: 26892526 Free PMC article. Review.
-
Anaortic techniques reduce neurological morbidity after off-pump coronary artery bypass surgery.Heart Lung Circ. 2008 Aug;17(4):299-304. doi: 10.1016/j.hlc.2007.11.138. Epub 2008 Feb 21. Heart Lung Circ. 2008. PMID: 18294911
-
Initial experience of an automated anastomotic distal device in off-pump CABG.Heart Surg Forum. 2004;7(5):E360-3. doi: 10.1532/HSF98.20041060. Heart Surg Forum. 2004. PMID: 15799903 Clinical Trial.
-
Evaluation of the enclose proximal anastomosis device in coronary artery bypass grafting.Ann Thorac Surg. 2005 Sep;80(3):1091-5. doi: 10.1016/j.athoracsur.2004.10.043. Ann Thorac Surg. 2005. PMID: 16122495 Clinical Trial.
-
[Heart positioners and proximal anastomotic devices in off-pump coronary artery bypass grafting].Kyobu Geka. 2009 Jul;62(8 Suppl):672-6. Kyobu Geka. 2009. PMID: 20715690 Review. Japanese.
Cited by
-
Cerebrovascular Events After No-Touch Off-Pump Coronary Artery Bypass Grafting, Conventional Side-Clamp Off-Pump Coronary Artery Bypass, and Proximal Anastomotic Devices: A Meta-Analysis.J Am Heart Assoc. 2016 Feb 18;5(2):e002802. doi: 10.1161/JAHA.115.002802. J Am Heart Assoc. 2016. PMID: 26892526 Free PMC article. Review.
-
The Use of Enclose(®)II Anastomosis Assist Device for the Proximal Coronary Branch Anastomosis to Vascular Graft.Ann Vasc Dis. 2010;3(1):84-6. doi: 10.3400/avd.AVDhdi08023. Epub 2010 Jul 21. Ann Vasc Dis. 2010. PMID: 23555395 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical