Comparative Study
. 2006 Feb;16(1):1-7.
doi: 10.1055/s-2006-923904.

The programmable adult Codman Hakim valve is useful even in very small children with hydrocephalus. A 7-year retrospective study with special focus on cost/benefit analysis

Affiliations
Comparative Study

The programmable adult Codman Hakim valve is useful even in very small children with hydrocephalus. A 7-year retrospective study with special focus on cost/benefit analysis

K Arnell et al. Eur J Pediatr Surg. 2006 Feb.

Abstract

Introduction: Ventriculo-peritoneal shunting is the most commonly used method for the treatment of paediatric hydrocephalus. The programmable valve with the ability to adjust the opening pressure non-invasively has made it easier to find exactly the right opening pressure for each child and reduce the risk of over- or under-drainage. The aim of this investigation was to study our clinical experience with the adult Codman Hakim programmable valve in children, with reference to complications and economic impact.

Methods: A seven-year retrospective study of 122 hydrocephalic children (aged from children born prematurely to 15 years old) shunted with the adult Codman Hakim programmable valve was performed.

Results: The programmable valve was the first shunt in 76 children and in 14 after prior ventricular drainage. The remaining 46 had different non-programmable systems as their first shunt. The most common reason for changing to a programmable valve was over-drainage. With the programmable valve, catheter-related complications, e.g. proximal (36%) or distal obstructions (30%), were the main reasons for surgical revision. Non-invasive pressure adjustment was performed in 73% of the children. Among the children with the programmable valve as their first shunt, 57 (75%) were adjusted, 12 (21%) had severe symptoms of over-drainage and would have required urgent surgical change of the valve if it had not been adjustable. A resetting of the opening pressure after MRI was found in 38% and accidental resetting occurred in 4%. Programmable valves are about twice as expensive as non-programmable valves. We estimated the increased cost of the valve and compared it with the savings from a reduction in the number of re-operations. The total cost for the programmable valve (as the primary shunt) in our study was less than that for expected re-operations due to over- or under-drainage when using non-programmable valves.

Conclusion: The programmable valve was easy to handle; only one size was required and the adjustment made it possible to achieve an optimal intraventricular pressure with a lower total cost, reduced hospital stay as well as an increased quality of life for the children.