. 2006 May;23(5):619-27.
doi: 10.1002/jmri.20554.

Image-guided and -monitored renal artery stenting using only MRI

Affiliations

Image-guided and -monitored renal artery stenting using only MRI

Daniel R Elgort et al. J Magn Reson Imaging. 2006 May.

Abstract

Purpose: To demonstrate the ability of a unique interventional MR system to be used safely and effectively as the only imaging modality for all phases of MR-guided stent-supported angioplasty.

Materials and methods: An experimental disease model of renal stenosis was created in six pigs. An interventional MR system, which employed previously reported tools for real-time catheter tracking with automated scan-plane positioning, adaptive image parameters, and radial true-FISP imaging with steady-state precession (True-FISP) imaging coupled with a high-speed reconstruction technique, was then used to guide all phases of the intervention, including: guidewire and catheter insertion, stent deployment, and confirmation of therapeutic success. Pre- and postprocedural X-ray imaging was used as a gold standard to validate the experimental results.

Results: All of the stent-supported angioplasty interventions were a technical success and were performed without complications. The average postoperative residual stenosis was 14.9%. The image guidance enabled the stents to be deployed with an accuracy of 0.98 +/- 0.69 mm. Additionally, using this interventional MRI system to guide renal artery stenting significantly reduces the procedure time, as compared to using X-ray fluoroscopy.

Conclusion: This study has clearly demonstrated the first successful treatment of renal artery stenting in an experimental animal model solely under MRI guidance and monitoring.