Electrocautery-induced fire during adenotonsillectomy: report of two cases
- PMID: 16563331
- DOI: 10.1016/j.jclinane.2005.09.032
Electrocautery-induced fire during adenotonsillectomy: report of two cases
Abstract
We describe two cases of flash fires in the oropharynx, secondary to electrocautery during adenotonsillectomies. We believe that in both cases, the leak around the uncuffed endotracheal tubes raised the oxygen concentration in the oropharynx. Cuffed endotracheal tubes provide many advantages, and their use should strongly be considered during adenotonsillectomy in children when electrocautery is to be used.
Comment in
-
Anesthesiologists must do a better job of preventing operating room fires.J Clin Anesth. 2006 Mar;18(2):81-2. doi: 10.1016/j.jclinane.2005.12.003. J Clin Anesth. 2006. PMID: 16563322 No abstract available.
Similar articles
-
Effect of cuffed and uncuffed endotracheal tubes on the oropharyngeal oxygen and volatile anesthetic agent concentration in children.Int J Pediatr Otorhinolaryngol. 2012 Jun;76(6):842-4. doi: 10.1016/j.ijporl.2012.02.055. Epub 2012 Mar 23. Int J Pediatr Otorhinolaryngol. 2012. PMID: 22444738
-
The prevention of fire during oropharyngeal electrosurgery.Ear Nose Throat J. 2003 Feb;82(2):107-9. Ear Nose Throat J. 2003. PMID: 12619467
-
[Airway fire during electrocautery dissection of the tonsils].Rev Esp Anestesiol Reanim. 2005 Dec;52(10):637-8. Rev Esp Anestesiol Reanim. 2005. PMID: 16435621 Spanish. No abstract available.
-
Fire safety in the operating room.Curr Opin Anaesthesiol. 2008 Dec;21(6):790-5. doi: 10.1097/ACO.0b013e328318693a. Curr Opin Anaesthesiol. 2008. PMID: 18997531 Review.
-
[Characteristics of anesthesiological care in adenotonsillectomy in children].Anesteziol Reanimatol. 1999 Jul-Aug;(4):59-62. Anesteziol Reanimatol. 1999. PMID: 10584374 Review. Russian. No abstract available.
Cited by
-
Crisis in the operating room: fires, explosions and electrical accidents.J Artif Organs. 2010 Sep;13(3):129-33. doi: 10.1007/s10047-010-0513-0. Epub 2010 Aug 14. J Artif Organs. 2010. PMID: 20711622 Review.
-
Management of a fire in the operating room.J Anesth. 2014 Apr;28(2):279-87. doi: 10.1007/s00540-013-1705-6. Epub 2013 Aug 30. J Anesth. 2014. PMID: 23989633 Review.
-
Factors involved in dental surgery fires: a review of the literature.Anesth Prog. 2014 Spring;61(1):21-5. doi: 10.2344/0003-3006-61.1.21. Anesth Prog. 2014. PMID: 24697822 Free PMC article. Review.
-
Options for Intraoperative Repair of a Cut Pilot Balloon on the Endotracheal Tube.Med Devices (Auckl). 2021 Sep 3;14:265-269. doi: 10.2147/MDER.S323982. eCollection 2021. Med Devices (Auckl). 2021. PMID: 34512044 Free PMC article.
-
Transoral Laser Microlaryngeal Surgery with MLT Tubes: A Retrospective Case Series.Indian J Otolaryngol Head Neck Surg. 2024 Oct;76(5):3922-3929. doi: 10.1007/s12070-024-04743-3. Epub 2024 Jun 6. Indian J Otolaryngol Head Neck Surg. 2024. PMID: 39376353