Abstract
Objective: To determine the surgery-related and hardware-related complications of deep-brain stimulation (DBS) at a single centre.
Methods: 262 consecutive patients (472 electrodes) operated for DBS in our department from February 1996 to March 2003 were retrospectively analysed to document acute adverse events (30 days postoperatively). The data of 180 of these patients were additionally revised to assess long-term complications (352 electrodes, mean follow-up 36.3 (SD 20.8) months).
Results: The frequency of minor intraoperative complications was 4.2% (11/262 patients). Transient (0.2%) or permanent (0.4%) neurological deficits, and in one case asymptomatic intracranial haemorrhage (0.2%), were registered as acute severe adverse events caused by surgery. Among minor acute complications were subcutaneous bleeding along the extension wire (1.2%) and haematoma at the pulse generator implantation site (1.2%). Skin infection caused by the implanted material was registered in 15 of 262 patients (5.7%). The infection rate during the first observation period was 1.5% (4/262 patients) and the late infection rate was 6.1% (11/180 patients). Partial or complete removal of the stimulation system was necessitated in 12 of 262 (4.6%) patients because of skin infection. During the long-term observation period, hardware-related problems were registered in 25 of 180 (13.9%) patients.
Conclusions: Stereotactic implantation of electrodes for DBS, if performed with multiplanar three-dimensional imaging and advanced treatment planning software, is a safe procedure with no mortality and low morbidity. The main causes for the patients' prolonged hospital stay and repeated surgery were wound infections and hardware-related complications.
Conflict of interest statement
Competing interests: YW was supported by a grant provided by Medtronic GmbH, Germany.
Similar articles
-
Surgical, medical, and hardware adverse events in a series of 141 patients undergoing subthalamic deep brain stimulation for Parkinson disease.World Neurosurg. 2010 Apr;73(4):338-44. doi: 10.1016/j.wneu.2010.01.017. World Neurosurg. 2010. PMID: 20849789
-
Hardware-related complications of deep brain stimulation: a ten year experience.Acta Neurochir (Wien). 2005 Oct;147(10):1061-4; discussion 1064. doi: 10.1007/s00701-005-0576-5. Epub 2005 Jul 25. Acta Neurochir (Wien). 2005. PMID: 16041470
-
Long-term surgical and hardware-related complications of deep brain stimulation.Stereotact Funct Neurosurg. 2011;89(2):89-95. doi: 10.1159/000323372. Epub 2011 Feb 2. Stereotact Funct Neurosurg. 2011. PMID: 21293168
-
Towards unambiguous reporting of complications related to deep brain stimulation surgery: A retrospective single-center analysis and systematic review of the literature.PLoS One. 2018 Aug 2;13(8):e0198529. doi: 10.1371/journal.pone.0198529. eCollection 2018. PLoS One. 2018. PMID: 30071021 Free PMC article.
-
Surgical and hardware complications of deep brain stimulation. A seven-year experience and review of the literature.Acta Neurochir (Wien). 2010 Dec;152(12):2053-62. doi: 10.1007/s00701-010-0749-8. Epub 2010 Jul 25. Acta Neurochir (Wien). 2010. PMID: 20658301 Review.
Cited by
-
An update on advanced therapies for Parkinson's disease: From gene therapy to neuromodulation.Front Surg. 2022 Sep 23;9:863921. doi: 10.3389/fsurg.2022.863921. eCollection 2022. Front Surg. 2022. PMID: 36211256 Free PMC article. Review.
-
Deep brain stimulation, lesioning, focused ultrasound: update on utility.AIMS Neurosci. 2023 Apr 26;10(2):87-108. doi: 10.3934/Neuroscience.2023007. eCollection 2023. AIMS Neurosci. 2023. PMID: 37426775 Free PMC article. Review.
-
Stimulation Efficiency With Decaying Exponential Waveforms in a Wirelessly Powered Switched-Capacitor Discharge Stimulation System.IEEE Trans Biomed Eng. 2018 May;65(5):1095-1106. doi: 10.1109/TBME.2017.2741107. Epub 2017 Aug 17. IEEE Trans Biomed Eng. 2018. PMID: 28829301 Free PMC article.
-
Impact of advancing age on post-operative complications of deep brain stimulation surgery for essential tremor.J Clin Neurosci. 2015 May;22(5):872-6. doi: 10.1016/j.jocn.2014.11.005. Epub 2015 Feb 7. J Clin Neurosci. 2015. PMID: 25669119 Free PMC article.
-
Single-Stage Deep Brain Stimulator Placement for Movement Disorders: A Case Series.Brain Sci. 2021 May 3;11(5):592. doi: 10.3390/brainsci11050592. Brain Sci. 2021. PMID: 34063572 Free PMC article.
References
-
- Gross R E. Deep brain stimulation in the treatment of neurological and psychiatric disease. Expert Rev Neurother 20044465–478. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical