Evaluation of the Dako IDEIA norovirus EIA assay for detection of norovirus using faecal specimens from Australian gastroenteritis outbreaks
- PMID: 16581657
- DOI: 10.1080/00313020600559645
Evaluation of the Dako IDEIA norovirus EIA assay for detection of norovirus using faecal specimens from Australian gastroenteritis outbreaks
Abstract
Aims: New techniques for detection of norovirus, a major cause of gastroenteritis, require ongoing evaluation. The aim of this study was to use material from gastroenteritis outbreaks in Victoria, Australia, to evaluate the sensitivity and specificity of the Dako IDEIA norovirus EIA assay, using both photometric and visual analysis.
Methods: A total of 130 faecal specimens from 41 gastroenteritis outbreaks were tested for norovirus by electron microscopy (EM), a two-round multiplex reverse transcription-polymerase chain reaction (RT-PCR) method and the IDEIA norovirus assay. All specimens with sufficient amplified product were sequenced to determine their norovirus genotype. In addition, six well-established RT-PCR protocols were used to test four EIA-positive, multiplex RT-PCR/EM-negative specimens. Also, a range of RT-PCR protocols was used to test a specimen positive for GII only by the multiplex RT-PCR but positive for GI and GII by the EIA. The effect of multiple freezing-thawing cycles on EIA positivity was tested on seven additional specimens. A further seven specimens, known to contain the gastroenteritis viruses sapovirus, adenovirus, astrovirus and rotavirus were also tested by the IDEIA norovirus assay.
Results: The IDEIA norovirus assay gave a single-specimen sensitivity and specificity of 66% and 85%, respectively (visual analysis compared with the multiplex RT-PCR), 63% and 88% (photometric analysis compared with the multiplex RT-PCR), 65% and 87% (visual analysis compared with the multiplex RT-PCR and/or EM) and 62% and 90% (photometric analysis compared with the multiplex RT-PCR and/or EM). None of the four EIA-positive specimens negative by the multiplex RT-PCR and/or EM was positive by any of the six alternative RT-PCR protocols. The specimen positive for GI and GII by EIA but for GII only by the multiplex RT-PCR was not positive for GI by any of the alternative RT-PCR protocols. A minimum of three specimens per outbreak had to be tested by the EIA to ensure that norovirus-positive outbreaks (multiplex RT-PCR and/or EM) were classified as positive for norovirus by the IDEIA norovirus assay (visual or photometric analysis). However, one specimen from a norovirus-negative outbreak (multiplex RT-PCR and/or EM) for which four specimens were provided was positive for norovirus by the IDEIA norovirus assay. Seven norovirus genotypes were identified by open reading frame 1 sequencing analysis and specimens from all seven norovirus genotypes (as well as an EM-positive/multiplex RT-PCR-negative specimen) were detected by the IDEIA norovirus assay by both visual and photometric analysis. Repeated freezing-thawing cycles (up to six) for faecal specimens did not reduce the sensitivity of the EIA assay but could render an EIA-negative specimen EIA-positive. The specimens positive for sapovirus, adenovirus, astrovirus and rotavirus were EIA-negative.
Conclusions: The IDEIA norovirus assay lacks the sensitivity and specificity to ascribe a particular result to a particular specimen, but could be useful for detecting norovirus in a gastroenteritis outbreak where specimens are plentiful, although it is difficult to avoid a risk of false positives. Since visual analysis can be used for result assessment almost as reliably as photometric analysis, the test kit would be useful for laboratories lacking specialist equipment such as a photometric microplate reader.
Similar articles
-
Multiplex real time RT-PCR for the detection and quantitation of norovirus genogroups I and II in patients with acute gastroenteritis.J Clin Virol. 2005 Jun;33(2):168-71. doi: 10.1016/j.jcv.2004.12.014. J Clin Virol. 2005. PMID: 15911433
-
A simple and rapid single-step multiplex RT-PCR to detect Norovirus, Astrovirus and Adenovirus in clinical stool samples.J Virol Methods. 2004 Jun 1;118(1):49-59. doi: 10.1016/j.jviromet.2004.01.016. J Virol Methods. 2004. PMID: 15158068
-
Sensitive and rapid detection of norovirus using duplex TaqMan reverse transcription-polymerase chain reaction.J Med Virol. 2008 May;80(5):913-20. doi: 10.1002/jmv.21142. J Med Virol. 2008. PMID: 18360905
-
[Diagnostic tests: Norovirus].Nihon Rinsho. 2005 Jul;63 Suppl 7:332-5. Nihon Rinsho. 2005. PMID: 16111266 Review. Japanese. No abstract available.
-
Laboratory diagnosis of noroviruses: present and future.Clin Lab Med. 2015 Jun;35(2):345-62. doi: 10.1016/j.cll.2015.02.008. Epub 2015 Apr 4. Clin Lab Med. 2015. PMID: 26004647 Review.
Cited by
-
Detection of norovirus antigens from recombinant virus-like particles and stool samples by a commercial norovirus enzyme-linked immunosorbent assay kit.J Clin Microbiol. 2006 Oct;44(10):3784-6. doi: 10.1128/JCM.01373-06. J Clin Microbiol. 2006. PMID: 17021111 Free PMC article.
-
Diagnostic accuracy and analytical sensitivity of IDEIA Norovirus assay for routine screening of human norovirus.J Clin Microbiol. 2010 Aug;48(8):2770-8. doi: 10.1128/JCM.00654-10. Epub 2010 Jun 16. J Clin Microbiol. 2010. PMID: 20554813 Free PMC article.
-
Norovirus.Clin Microbiol Rev. 2015 Jan;28(1):134-64. doi: 10.1128/CMR.00075-14. Clin Microbiol Rev. 2015. PMID: 25567225 Free PMC article. Review.
-
Application of a modified loop-mediated isothermal amplification kit for detecting Norovirus genogroups I and II.J Med Virol. 2009 Dec;81(12):2072-8. doi: 10.1002/jmv.21626. J Med Virol. 2009. PMID: 19856470 Free PMC article.
-
European multicenter evaluation of commercial enzyme immunoassays for detecting norovirus antigen in fecal samples.Clin Vaccine Immunol. 2007 Oct;14(10):1349-55. doi: 10.1128/CVI.00214-07. Epub 2007 Aug 22. Clin Vaccine Immunol. 2007. PMID: 17715333 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical