Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI
- PMID: 16595498
Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI
Abstract
PET combined with CT has proven to be a valuable multimodality imaging device revealing both functional and anatomic information. Although PET/CT has become completely integrated into routine clinical application and also has been used in small-animal imaging, CT provides only limited soft-tissue contrast and, in preclinical studies, exposes the animal to a relatively high radiation dose. Unlike CT, MRI provides good soft-tissue contrast even without application of contrast agents and, furthermore, does not require ionizing radiation.
Methods: This project focused on combining a high-resolution PET scanner with a 7-T MRI system for animal research. Because classic PET detectors based on photomultiplier tubes cannot be used in high magnetic fields, we used a detector technology based on 10 x 10 lutetium oxyorthosilicate crystal arrays and 3 x 3 avalanche photodiode arrays. A ring of such PET detectors will ultimately be used as an insert for the 119-mm-diameter MRI bore.
Results: Initial measurements with 1 PET detector module in the 7-T field during application of MRI sequences were encouraging. Position profiles from the PET detectors and a first MR image of a mouse could be acquired simultaneously.
Conclusion: Further work will concentrate on the construction of a full PET detector ring with compact, integrated electronics.
Similar articles
-
Multimodal imaging approaches: PET/CT and PET/MRI.Handb Exp Pharmacol. 2008;(185 Pt 1):109-32. doi: 10.1007/978-3-540-72718-7_6. Handb Exp Pharmacol. 2008. PMID: 18626801 Review.
-
Initial results of simultaneous PET/MRI experiments with an MRI-compatible silicon photomultiplier PET scanner.J Nucl Med. 2012 Apr;53(4):608-14. doi: 10.2967/jnumed.111.097501. Epub 2012 Mar 13. J Nucl Med. 2012. PMID: 22414638
-
Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard.J Nucl Med. 2005 Dec;46(12):2083-91. J Nucl Med. 2005. PMID: 16330574
-
PET/MR images acquired with a compact MR-compatible PET detector in a 7-T magnet.Radiology. 2007 Sep;244(3):807-14. doi: 10.1148/radiol.2443061756. Radiology. 2007. PMID: 17709830
-
Whole-body magnetic resonance imaging and positron emission tomography-computed tomography in oncology.Top Magn Reson Imaging. 2007 Jun;18(3):193-202. doi: 10.1097/RMR.0b013e318093e6bo. Top Magn Reson Imaging. 2007. PMID: 17762383 Review.
Cited by
-
Split gradient coils for simultaneous PET-MRI.Magn Reson Med. 2009 Nov;62(5):1106-11. doi: 10.1002/mrm.22143. Magn Reson Med. 2009. PMID: 19780167 Free PMC article.
-
In vivo Noninvasive Small Animal Molecular Imaging.Osong Public Health Res Perspect. 2012 Mar;3(1):48-59. doi: 10.1016/j.phrp.2012.02.002. Osong Public Health Res Perspect. 2012. PMID: 24159487 Free PMC article. Review.
-
Automatic 3D registration of dynamic stress and rest (82)Rb and flurpiridaz F 18 myocardial perfusion PET data for patient motion detection and correction.Med Phys. 2011 Nov;38(11):6313-26. doi: 10.1118/1.3656951. Med Phys. 2011. PMID: 22047396 Free PMC article.
-
A look ahead: PET/MR versus PET/CT.Eur J Nucl Med Mol Imaging. 2009 Mar;36 Suppl 1:S3-9. doi: 10.1007/s00259-008-0940-9. Eur J Nucl Med Mol Imaging. 2009. PMID: 19104804
-
Molecular imaging in oncology: the acceptance of PET/CT and the emergence of MR/PET imaging.Eur Radiol. 2011 Mar;21(3):548-54. doi: 10.1007/s00330-010-2033-y. Epub 2010 Dec 21. Eur Radiol. 2011. PMID: 21174096 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical