BiliCheck transcutaneous bilirubinometer: a screening tool for neonatal jaundice in the Chinese population
- PMID: 16603775
BiliCheck transcutaneous bilirubinometer: a screening tool for neonatal jaundice in the Chinese population
Abstract
Objective: To verify the usefulness of the BiliCheck transcutaneous bilirubin meter as a screening device for neonatal jaundice in a Chinese population compared with the Minolta bilirubin meter.
Design: A prospective correlation study that compared transcutaneous bilirubin measurements with serum bilirubin levels.
Setting: Obstetric ward and a neonatal unit of a regional hospital in Hong Kong.
Patients: Neonates with gestation above 32 weeks with neonatal jaundice who were admitted between April 2001 and February 2002.
Main outcome measures: Transcutaneous measurements of serum bilirubin obtained from the forehead and the sternum with two instruments: BiliCheck and Minolta Airshields JM 102.
Results: A total of 77 term and six near-term babies (gestation, 32-37 weeks) were recruited. The mean age at the time of data collection was 3.96 days (range, 2-9 days). The correlations between serum bilirubin and transcutaneous bilirubin measurements of the two devices at the two sites were high, with a coefficient of 0.718 (95% confidence interval, 0.610-0.800; n=100) for forehead measurements, and 0.814 (95% confidence interval, 0.740-0.870; n=99) for sternum using the Minolta Airshields JM 102; and a coefficient of 0.757 (95% confidence interval, 0.657-0.827; n=98) for forehead measurements, and 0.794 (95% confidence interval, 0.700-0.862; n=92) for sternum using the BiliCheck. For BiliCheck, a cut-off point of 250 micromol/L at the forehead and 260 micromol/L at the sternum had a specificity of 61.9% and 70.0%, respectively with a sensitivity of 100% for the detection of serum bilirubin concentrations of 250 micromol/L or higher. This level is commonly used as the level for initiation of treatment such as phototherapy.
Conclusion: BiliCheck is a useful screening tool for neonatal jaundice in the Chinese population and is comparable with the Minolta Airshields JM 102.
Similar articles
-
Comparison of two transcutaneous bilirubinometers--Minolta AirShields Jaundice Meter JM103 and Spectrx Bilicheck--in Thai neonates.Southeast Asian J Trop Med Public Health. 2005 Nov;36(6):1533-7. Southeast Asian J Trop Med Public Health. 2005. PMID: 16610658
-
Evaluation of a point-of-care transcutaneous bilirubinometer in Chinese neonates at an accident and emergency department.Hong Kong Med J. 2008 Oct;14(5):356-60. Hong Kong Med J. 2008. PMID: 18840905
-
Improved transcutaneous bilirubinometry: comparison of SpectR(X) BiliCheck and Minolta Jaundice Meter JM-102 for estimating total serum bilirubin in a normal newborn population.J Perinatol. 2002 Jan;22(1):12-4. doi: 10.1038/sj.jp.7210592. J Perinatol. 2002. PMID: 11840236 Clinical Trial.
-
Non-invasive bilirubinometry in neonatal jaundice.Semin Neonatol. 2002 Apr;7(2):129-33. doi: 10.1053/siny.2002.0100. Semin Neonatol. 2002. PMID: 12208097 Review.
-
Transcutaneous bilirubinometry: its role in the assessment of neonatal jaundice.Clin Biochem. 1997 Feb;30(1):1-9. doi: 10.1016/s0009-9120(96)00131-2. Clin Biochem. 1997. PMID: 9056103 Review.
Cited by
-
Transcutaneous bilirubin nomogram for predicting neonatal hyperbilirubinemia in healthy term and late-preterm Chinese infants.Eur J Pediatr. 2011 Feb;170(2):185-91. doi: 10.1007/s00431-010-1281-9. Epub 2010 Sep 3. Eur J Pediatr. 2011. PMID: 20814696
-
The value of bilicheck® as a screening tool for neonatal jaundice in the South of iran.Iran J Med Sci. 2013 Jun;38(2):122-8. Iran J Med Sci. 2013. PMID: 23825892 Free PMC article.
-
Reliability of transcutaneous bilirubin determination based on skin color determined by a neonatal skin color scale of our own.Eur J Pediatr. 2021 Feb;180(2):607-616. doi: 10.1007/s00431-020-03885-0. Epub 2021 Jan 6. Eur J Pediatr. 2021. PMID: 33409587
-
Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates.Eur J Pediatr. 2014 Feb;173(2):173-9. doi: 10.1007/s00431-013-2142-0. Epub 2013 Aug 17. Eur J Pediatr. 2014. PMID: 23955487
-
Demystifying non-invasive approaches for screening jaundice in low resource settings: a review.Front Pediatr. 2023 Nov 20;11:1292678. doi: 10.3389/fped.2023.1292678. eCollection 2023. Front Pediatr. 2023. PMID: 38054187 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous