Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry
- PMID: 16606871
- DOI: 10.1001/archopht.124.4.471
Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry
Abstract
Objective: To evaluate the effect of central corneal thickness (CCT), corneal curvature, and axial length on applanation tonometry in an in vivo study.
Methods: In a masked, prospective clinical trial, we examined 125 eyes of 125 patients scheduled for cataract surgery. Corneal curvature was measured by means of keratometry and axial length by A-scan ultrasonography. By cannulating the anterior chamber before surgery, intraocular pressure (IOP) was set to 20, 35, and 50 mm Hg in a closed system by means of a water column. After measuring thickness, the IOP was measured with an applanation tonometer. Pearson product moment correlations and multiple linear regression analyses were performed, and significance levels were evaluated by the paired, 2-tailed t test.
Results: The difference between measured and real IOP was significantly dependent (P < .001) on CCT. The associations between IOP and corneal curvature or IOP and axial length were not statistically significant (P = .31). The association between IOP reading and CCT is shown in the "Dresdner correction table," which illustrates an approximately 1-mm Hg correction for every 25-microm deviation from a CCT of 550 microm. The correction values were positive as thickness decreased and negative as thickness increased.
Conclusions: Central corneal thickness significantly affects IOP readings obtained by applanation tonometry according to the Goldmann principle. A correction of IOP readings by considering CCT according to the Dresdner correction table might be helpful for determining an accurate IOP value.
Similar articles
-
Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonometry.Ophthalmology. 2007 Jan;114(1):20-6. doi: 10.1016/j.ophtha.2006.06.047. Epub 2006 Oct 27. Ophthalmology. 2007. PMID: 17070592
-
Effects of central corneal thickness, central corneal power, and axial length on intraocular pressure measurement assessed with goldmann applanation tonometry.Jpn J Ophthalmol. 2008 Sep-Oct;52(5):353-356. doi: 10.1007/s10384-008-0535-9. Epub 2008 Nov 11. Jpn J Ophthalmol. 2008. PMID: 18991034
-
Effects of central corneal thickness and corneal curvature on the intraocular pressure measurement by Goldmann applanation tonometer and ocular blood flow pneumatonometer.Clin Exp Ophthalmol. 2006 Aug;34(6):516-20. doi: 10.1111/j.1442-9071.2006.01266.x. Clin Exp Ophthalmol. 2006. PMID: 16925697 Clinical Trial.
-
Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach.Surv Ophthalmol. 2000 Mar-Apr;44(5):367-408. doi: 10.1016/s0039-6257(00)00110-7. Surv Ophthalmol. 2000. PMID: 10734239 Review.
-
Finite element simulation of Goldmann tonometry after refractive surgery.Clin Biomech (Bristol). 2020 Jan;71:24-28. doi: 10.1016/j.clinbiomech.2019.09.007. Epub 2019 Oct 5. Clin Biomech (Bristol). 2020. PMID: 31677547
Cited by
-
Validity of Tono-pachymetry for Measuring Corrected Intraocular Pressure in Non-surgical and Post-photorefractive Keratectomy Eyes.Korean J Ophthalmol. 2017 Feb;31(1):44-51. doi: 10.3341/kjo.2017.31.1.44. Epub 2017 Feb 2. Korean J Ophthalmol. 2017. PMID: 28243023 Free PMC article.
-
Association analysis of cigarette smoking with onset of primary open-angle glaucoma and glaucoma-related biometric parameters.BMC Ophthalmol. 2012 Nov 27;12:59. doi: 10.1186/1471-2415-12-59. BMC Ophthalmol. 2012. PMID: 23186177 Free PMC article.
-
The effects of the modulus of the lens material on intraocular pressure measurement through soft contact lenses.Ir J Med Sci. 2013 Sep;182(3):331-5. doi: 10.1007/s11845-012-0881-y. Epub 2012 Dec 4. Ir J Med Sci. 2013. PMID: 23207914
-
Intraocular Pressure (IOP) in Patients with Acromegaly versus Healthy Controls: A Systematic Review and Meta-Analysis.Vision (Basel). 2024 Sep 12;8(3):54. doi: 10.3390/vision8030054. Vision (Basel). 2024. PMID: 39311322 Free PMC article. Review.
-
Axial length acquisition success rates and agreement of two swept-source optical biometers in eyes with dense cataracts.Front Med (Lausanne). 2024 Sep 25;11:1449867. doi: 10.3389/fmed.2024.1449867. eCollection 2024. Front Med (Lausanne). 2024. PMID: 39386744 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical