Stented bioprosthetic valve hemodynamics: is the supra-annular implant better than the intra-annular?
- PMID: 16607907
Stented bioprosthetic valve hemodynamics: is the supra-annular implant better than the intra-annular?
Abstract
Background and aim of the study: The use of stented bioprostheses for aortic valve replacement (AVR) in elderly patients with a small aortic annulus may result in unsatisfactory hemodynamic performance of the prosthesis. To overcome this limitation, new bioprostheses have been designed for complete supra-annular implantation, but the actual hemodynamic advantage of the supra-annular implant over the intra-annular has not been fully investigated. Accordingly, the hemodynamic performance of the same stented bioprosthesis (except for sewing ring design) implanted in the supra-annular and conventional intra-annular seating was compared.
Methods: Twenty-two patients received an intra-annular implant, and 38 a supra-annular implant. Age (74 +/- 5 versus 76 +/- 5 years, p = 0.54), gender (55% versus 50% males, p = 0.79) and body surface area (1.74 +/- 0.2 versus 1.81 +/- 0.2 m2, p = 0.13) were similar in both subgroups, who underwent echocardiography at 8 +/- 2 and 6 +/- 2 months after surgery, respectively (p = 0.09).
Results: The two patient subgroups had similar preoperative left ventricular outflow tract diameters (2.06 +/- 0.2 and 2.1 +/- 0.2 cm; p = 0.62), average size of implanted prosthesis (21.0 and 21.3 mm; p = 0.44) and mean transprosthetic flow rate (246 +/- 70 and 218 +/- 58 ml/s; p = 0.12). Mean (8 +/- 3 and 19 +/- 8 mmHg, p < 0.0001), and peak (17 +/- 6 and 40 +/- 13 mmHg; p < 0.0001) transprosthetic gradients were lower, and mean effective orifice area (EOA) (1.78 +/- 0.4 and 1.45 +/- 0.5 cm2, p = 0.006) was higher in patients with supra-annular implants than in those with intraannular. The incidence of patient-prosthesis mismatch (EOA index < 0.85 cm2/m2) decreased from 50% to 34% (p < 0.0001), with no case of severe mismatch using the supra-annular implant. During follow up, a left ventricular mass reduction occurred in patients with supra-annular implants (from 225 +/- 110 to 173 +/- 59 g/m2; p < 0.03), but not in patients with intra-annular implants (173 +/- 62 and 186 +/- 64 g/m2; p = 0.87)
Conclusion: The study results showed that, compared to intra-annular implantation, supra-annular implantation of bioprosthetic stented valves in the aortic position was associated with a significantly better hemodynamic performance of the prosthesis and significant regression of left ventricular hypertrophy.
Similar articles
-
The Toronto root stentless valve in the subcoronary position is hemodynamically superior to the mosaic stented completely supra-annular bioprosthesis.J Heart Valve Dis. 2005 Nov;14(6):814-21; discussion 821. J Heart Valve Dis. 2005. PMID: 16359064
-
Results of aortic valve replacement with a new supra-annular pericardial stented bioprosthesis.Ann Thorac Surg. 2006 Dec;82(6):2133-8. doi: 10.1016/j.athoracsur.2006.07.064. Ann Thorac Surg. 2006. PMID: 17126124
-
Influence of completely supra-annular placement of bioprostheses on exercise hemodynamics in patients with a small aortic annulus.J Thorac Cardiovasc Surg. 2007 May;133(5):1234-41. doi: 10.1016/j.jtcvs.2006.10.074. J Thorac Cardiovasc Surg. 2007. PMID: 17467435
-
Clinical and hemodynamic implications of supra-annular implant of biological aortic valves.J Cardiovasc Med (Hagerstown). 2006 Jul;7(7):524-32. doi: 10.2459/01.JCM.0000234771.96324.28. J Cardiovasc Med (Hagerstown). 2006. PMID: 16801814 Review.
-
Biological aortic valve replacement: advantages and optimal indications of stentless compared to stented valve substitutes. A review.Gen Thorac Cardiovasc Surg. 2018 May;66(5):247-256. doi: 10.1007/s11748-018-0884-3. Epub 2018 Jan 10. Gen Thorac Cardiovasc Surg. 2018. PMID: 29322433 Review.
Cited by
-
Optimising the Haemodynamics of Aortic Valve-in-valve Procedures.Interv Cardiol. 2017 May;12(1):40-43. doi: 10.15420/icr.2016:25:2. Interv Cardiol. 2017. PMID: 29588729 Free PMC article.
-
Initial experience with the Crea Aortic Valve System - a first-in-human study.AsiaIntervention. 2025 Mar 20;11(1):52-59. doi: 10.4244/AIJ-D-24-00023. eCollection 2025 Mar. AsiaIntervention. 2025. PMID: 40114741 Free PMC article.
-
Aortic root widening: "pro et contra".Indian J Thorac Cardiovasc Surg. 2022 Apr;38(Suppl 1):91-100. doi: 10.1007/s12055-020-01125-6. Epub 2021 Feb 22. Indian J Thorac Cardiovasc Surg. 2022. PMID: 35463701 Free PMC article.
-
Determinants of effective orifice area in aortic valve replacement: anatomic and clinical factors.J Thorac Dis. 2020 May;12(5):1942-1951. doi: 10.21037/jtd-20-188. J Thorac Dis. 2020. PMID: 32642097 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Research Materials