Randomized Controlled Trial
. 2006 Mar;8(3):216-20.
doi: 10.1093/europace/euj027. Epub 2006 Feb 10.

Role of ventricular Autocapture function in increasing longevity of DDDR pacemakers: a prospective study

Affiliations
Randomized Controlled Trial

Role of ventricular Autocapture function in increasing longevity of DDDR pacemakers: a prospective study

Giuseppe Boriani et al. Europace. 2006 Mar.

Abstract

Aims: Autocapture is an algorithm for automatic adaptation of ventricular output to capture threshold. The aim of this prospective study was to estimate the effects of ventricular Autocapture algorithm on DDD-DDDR pacemaker longevity.

Methods and results: Eighty-three patients implanted with a DDD-DDDR pacemaker (Affinity or Entity; St Jude Medical, USA) were enrolled and the Autocapture function was activated pre-discharge. Ventricular pulse duration was randomly programmed at 0.3 or 0.4 ms, with a cross-over at 8-12 weeks and again at 13-14 months. Diagnostic data were retrieved from device memory and by calculating battery current drain from long-term threshold recordings; device longevity was estimated at the following settings: Autocapture with a pulse duration of 0.3 and 0.4 ms, respectively, standard output (3.5 V, 0.4 ms) and conventional low output programming (2.5 V, 0.4 ms). According to a series of assumptions, Autocapture was associated with a 55-60% increase in estimated device longevity compared with standard output programming and a 6-7% increase in longevity compared with low output programming. No significant differences were found between Autocapture programmed with a pulse duration of 0.3 or 0.4 ms. In projections to a 10-year follow-up, use of the Autocapture function resulted in a 42% reduction in pacing-related estimated costs compared with standard output programming at 3.5 V, 0.4 ms.

Conclusion: Pacing with constant adaptation of ventricular output in dual-chamber devices has the potential to increase generator longevity and to reduce sizeably pacing-related costs compared with standard programming.