. 2006 Sep;19(3):276-8.
doi: 10.1007/s10278-006-0265-9.

ROC analysis for diagnostic accuracy of fracture by using different monitors

Affiliations

ROC analysis for diagnostic accuracy of fracture by using different monitors

Zhigang Liang et al. J Digit Imaging. 2006 Sep.

Abstract

Background: The purpose of this study was to compare diagnostic accuracy by using two types of monitors.

Methods: Four radiologists with 10 years experience twice interpreted the films of 77 fracture cases by using the ViewSonic P75f+ and BARCO MGD221 monitors, with a time interval of 3 weeks. Each time the radiologists used one type of monitor to interpret the images. The image browser used was the Unisight software provided by Atlastiger Company (Shanghai, China), and interpretation result was analyzed via the LABMRMC software.

Results: In studies of receiver operating characteristics to score the presence or absence of fracture, the results of images interpreted through monochromic monitors showed significant statistical difference compared to those interpreted using the color monitors.

Conclusion: A significant difference was observed in the results obtained by using two kinds of monitors. Color monitors cannot serve as substitutes for monochromatic monitors in the process of interpreting computed radiography (CR) images with fractures.

PubMed Disclaimer

Figures

Fig 1
Fig 1
ROC curves for radiologist A by using two types of monitors.
Fig 2
Fig 2
ROC curves for radiologist B by using two types of monitors.
Fig 3
Fig 3
ROC curves for radiologist C by using two types of monitors.
Fig 4
Fig 4
ROC curves for radiologist D by using two types of monitors.

References

    1. Hertzberg BS, Kliewer MA, Paulson EK, et al. PACS in sonography: accuracy of interpretation using film compared with monitor display. Picture archiving and communication systems. AJR Am J Roentgenol. 1999;173:1175–1179. - PubMed
    1. Ishigaki T, Endo T, Ikeda M, et al. Subtle pulmonary disease: detection with computed radiography versus conventional chest radiography. Radiology. 1996;201:51–60. - PubMed
    1. Kondoh H, Hatazawa K, Kozuka T. The performance of general-purpose colour CRT monitors in PACS environment. Comput Methods Programs Biomed. 2001;7;66(1):63–68. doi: 10.1016/S0169-2607(01)00136-5. - DOI - PubMed
    1. Passadore DJ, Isoardi RA, Ariza PP, Padin C. Use of a low-cost, PC-based image review workstation at a radiology department. J Digit Imaging. 2001;6;14(2 Suppl 1):222–223. doi: 10.1007/BF03190346. - DOI - PMC - PubMed
    1. Partan G, Mayrhofer R, Urban M, et al. Diagnostic performance of liquid crystal and cathode-ray-tube monitors in brain computed tomography. Eur Radiol. 2003;13:2397–2401. doi: 10.1007/s00330-003-1822-y. - DOI - PubMed