Impedance cardiography: a potential monitor for hemodialysis
- PMID: 16631198
- DOI: 10.1016/j.jss.2006.03.004
Impedance cardiography: a potential monitor for hemodialysis
Abstract
Background: Impedance cardiography (ICG) technology has improved dramatically, and at least one device now can give a measurement of fluid status by using thoracic fluid content (TFC), along with cardiac output (CO) and cardiac index (CI). With a built-in sphygmomanometer cuff, it can also provide blood pressure (BP) and systemic vascular resistance index (SVRI). A currently available small portable ICG that provides reliable measures of fluid status could be an ideal noninvasive monitor for hemodialysis (HD), with the potential of helping avoid significant hemodynamic instability during HD.
Methods: A case series of patients with chronic renal failure was studied while undergoing HD using ICG (BioZ, CardioDynamics, Int. Corp., San Diego, CA). Parameters recorded at 15-min intervals included TFC, CI, BP (systolic, diastolic, and mean arterial), SVRI, and heart rate. Using the Pearson method, the percentage changes in each of the parameters during the HD session were correlated to the amount of fluid removed (FR), normalized to body weight.
Results: Forty-one patients were enrolled, but six patients were excluded due to incomplete data; therefore, 35 patients (13 men and 22 women) formed the basis of the analysis. The age range was 28 to 87 (mean 55.1 +/- 16.1) years. The amount of FR was 2.88 +/- 1.13 L (37.3 +/- 14.6 ml/kg). TFC decreased in all patients during the HD session (average reduction 12.7 +/- 8 kohms(-1)); whereas all other hemodynamic parameters showed both increases and decreases. The correlation of change in TFC with FR was moderate (r = 0.579, P = 0.0003); other hemodynamic parameters showed a poor correlation with FR. Neither the standard hemodynamic parameters nor the ICG device's special parameters were able to identify the five patients in this series who experienced significant hemodynamic instability or intradialytic hypotension.
Conclusion: TFC, measured easily and noninvasively using ICG, correlates with the amount of fluid removed during HD. In comparison with the other hemodynamic parameters measured, TFC changed most consistently with fluid removal. Whether or not serial TFC measurements in a given patient at different HD sessions can guide the extent of FR will require additional study. This compact, easily operated, and nonobtrusive ICG device with the capability for continuously providing the standard hemodynamic parameters plus CO, TFC, and standard limb lead electrocardiography could replace current monitoring systems.
Similar articles
-
Bioreactance: a new tool for cardiac output and thoracic fluid content monitoring during hemodialysis.Hemodial Int. 2009 Oct;13(4):512-7. doi: 10.1111/j.1542-4758.2009.00386.x. Epub 2009 Sep 16. Hemodial Int. 2009. PMID: 19758300
-
Prognostic value of hemodynamic findings from impedance cardiography in hypertensive stroke.Am J Hypertens. 2005 Feb;18(2 Pt 2):65S-72S. doi: 10.1016/j.amjhyper.2004.11.027. Am J Hypertens. 2005. PMID: 15752935
-
Systolic blood pressure does not reliably identify vasoactive status in chronic heart failure.Am J Hypertens. 2005 Feb;18(2 Pt 2):82S-86S. doi: 10.1016/j.amjhyper.2004.11.024. Am J Hypertens. 2005. PMID: 15752937
-
Transthoracic impedance cardiography: a noninvasive method of hemodynamic assessment.Heart Fail Clin. 2009 Apr;5(2):161-8. doi: 10.1016/j.hfc.2008.12.001. Heart Fail Clin. 2009. PMID: 19249685 Review.
-
Haemodialysis with on-line monitoring equipment: tools or toys?Nephrol Dial Transplant. 2005 Jan;20(1):22-33. doi: 10.1093/ndt/gfh555. Nephrol Dial Transplant. 2005. PMID: 15632348 Review.
Cited by
-
Use of Bioimpedance to Assess Changes in Hemodynamics During Acute Administration of CPAP.Cardiol Res. 2011 Apr;2(2):51-57. doi: 10.4021/cr18w. Epub 2011 Mar 25. Cardiol Res. 2011. PMID: 28348662 Free PMC article.
-
Evaluation of intradialytic hypotension using impedance cardiography.Int Urol Nephrol. 2011 Sep;43(3):855-64. doi: 10.1007/s11255-010-9746-3. Epub 2010 May 7. Int Urol Nephrol. 2011. PMID: 20449654 Clinical Trial.
-
A novel wearable bioimpedance sensor for continuous monitoring of fluid balance: a study on isotonic hypovolemia in healthy adults.J Clin Monit Comput. 2025 Apr;39(2):379-391. doi: 10.1007/s10877-024-01245-z. Epub 2024 Dec 4. J Clin Monit Comput. 2025. PMID: 39630397 Free PMC article. Clinical Trial.
-
Evaluation of heart function with impedance cardiography in acute myocardial infarction patients.Int J Clin Exp Med. 2014 Mar 15;7(3):719-27. eCollection 2014. Int J Clin Exp Med. 2014. PMID: 24753769 Free PMC article.
-
Alteration autonomic control of cardiac function during hemodialysis predict cardiovascular outcomes in end stage renal disease patients.Sci Rep. 2019 Dec 11;9(1):18783. doi: 10.1038/s41598-019-55001-4. Sci Rep. 2019. PMID: 31827106 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical