Stability of the EasyOne ultrasonic spirometer for use in general practice
- PMID: 16635089
- DOI: 10.1111/j.1440-1843.2006.00842.x
Stability of the EasyOne ultrasonic spirometer for use in general practice
Abstract
Objective and background: Spirometry is recommended for the diagnosis and management of chronic respiratory diseases in the community. Spirometer accuracy is critical, but few general practitioners meet the American Thoracic Society and European Respiratory Society (ATS/ERS) recommendation for daily calibration. The aim of this study was to assess the accuracy and stability of a portable ultrasonic spirometer (EasyOne) that the manufacturer claims does not require regular calibration.
Methods: Six EasyOne spirometers were used in a practice-based spirometry study. Inspiratory and expiratory accuracy was checked periodically using a certified 3-L syringe. Paired calibration checks were performed using a dedicated mouthpiece assembly (spirette) and randomly selected spirettes.
Results: The six spirometers were used for up to 26 weeks (mean 23.9 weeks) and a total of 1041 spirometry tests and 75 syringe calibrations were performed. All inspiratory and expiratory calibration checks using a dedicated or randomly selected spirette met the ATS/ERS accuracy criterion (3.00+/-0.105 L). The mean (range) expiratory volume deviation from target volume (3.00 L) was 0.011 L (-0.06-0.09 L) using a dedicated spirette and 0.046 L (-0.09-0.10 L) using randomly selected spirettes. The deviation from target was not affected by the mean flow generated during the calibration procedure. There was no change in calibration during the study.
Conclusions: This study supports the manufacturer's claim that the EasyOne spirometer maintains its calibration during routine clinical use in general practice and does not require daily calibration as specified in international spirometry guidelines.
Similar articles
-
Four-year calibration stability of the EasyOne portable spirometer.Respir Care. 2010 Jul;55(7):873-7. Respir Care. 2010. PMID: 20587099
-
Reproducibility and validity of a handheld spirometer.Respir Care. 2008 Apr;53(4):433-41. Respir Care. 2008. PMID: 18364054 Free PMC article.
-
The long-term stability of portable spirometers used in a multinational study of the prevalence of chronic obstructive pulmonary disease.Respir Care. 2006 Oct;51(10):1167-71. Respir Care. 2006. PMID: 17005063
-
Primary care spirometry.Eur Respir J. 2008 Jan;31(1):197-203. doi: 10.1183/09031936.00066607. Eur Respir J. 2008. PMID: 18166597 Review.
-
Science reflects history as society influences science: brief history of "race," "race correction," and the spirometer.Adv Physiol Educ. 2018 Jun 1;42(2):163-165. doi: 10.1152/advan.00196.2017. Adv Physiol Educ. 2018. PMID: 29616572 Review.
Cited by
-
Decreased Pulmonary Function Over 5 Years in US Firefighters.J Occup Environ Med. 2020 Oct;62(10):816-819. doi: 10.1097/JOM.0000000000001944. J Occup Environ Med. 2020. PMID: 32590426 Free PMC article.
-
COPD and cognitive impairment: the role of hypoxemia and oxygen therapy.Int J Chron Obstruct Pulmon Dis. 2010 Sep 7;5:263-9. doi: 10.2147/copd.s10684. Int J Chron Obstruct Pulmon Dis. 2010. PMID: 20856825 Free PMC article.
-
Disturbed sleep among COPD patients is longitudinally associated with mortality and adverse COPD outcomes.Sleep Med. 2012 May;13(5):476-83. doi: 10.1016/j.sleep.2011.12.007. Epub 2012 Mar 18. Sleep Med. 2012. PMID: 22429651 Free PMC article.
-
Cardiac autonomic functioning is impaired among allogeneic hematopoietic stem cell transplantation survivors: a controlled study.Bone Marrow Transplant. 2017 Jan;52(1):66-72. doi: 10.1038/bmt.2016.176. Epub 2016 Jul 4. Bone Marrow Transplant. 2017. PMID: 27376446 Clinical Trial.
-
Influence of anxiety on health outcomes in COPD.Thorax. 2010 Mar;65(3):229-34. doi: 10.1136/thx.2009.126201. Thorax. 2010. PMID: 20335292 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous