Calibration of Infra-red CO(2) Gas Analyzers
- PMID: 16657177
- PMCID: PMC396226
- DOI: 10.1104/pp.44.8.1122
Calibration of Infra-red CO(2) Gas Analyzers
Abstract
Precision gas mixing pumps produce CO(2) gas mixtures for the calibration of infra-red CO(2) gas analyzers equivalent in accuracy to the standard CO(2) gas mixtures (+/- 1%) supplied by the National Bureau of Standards, Washington, D. C.The calibration of infra-red gas analyzers by the pressure difference method and by concentration differences did not agree. A factor of 0.709 was necessary to correct the pressure difference method for the effects of pressure broadening with our analyzer.
References
-
- J Sci Instrum. 1968 Oct;1(10):1003-5 - PubMed