Sensitivity of the Multispot HIV-1/HIV-2 rapid test using samples from human immunodeficiency virus type 1-positive individuals with various levels of exposure to highly active antiretroviral therapy
- PMID: 16672414
- PMCID: PMC1479184
- DOI: 10.1128/JCM.44.5.1831-1833.2006
Sensitivity of the Multispot HIV-1/HIV-2 rapid test using samples from human immunodeficiency virus type 1-positive individuals with various levels of exposure to highly active antiretroviral therapy
Abstract
The Multispot HIV-1/HIV-2 rapid test detects human immunodeficiency virus type 1 (HIV-1) gp41 antibodies, which can wane over time in some HIV-1-infected populations, resulting in false-negative screening results. Multispot sensitivity was 100% using 248 sera from one such population, and it correctly identified serostatus in individuals who previously tested false negative with rapid testing.
References
-
- Bio-Rad Laboratories. 2004. Multispot HIV-1/HIV-2 Rapid Test package insert. Bio-Rad Laboratories, Redmond, Wash.
-
- Calero, E., J. Malia, R. Sawyer, L. Brown, D. Birx, B. Branson, and N. Michael. 2002. Rapid HIV-1 diagnostic algorithms for use in HIV infection screening. Presented at the XIV International AIDS Conference, Barcelona, Spain.
-
- Holguin, A. 2004. Evaluation of three rapid tests for detection of antibodies to HIV-1 non-B subtypes. J. Virol. Methods 115:105-107. - PubMed
-
- Kim, J. H., J. R. Mascola, S. Ratto-Kim, T. C. VanCott, L. Loomis-Price, J. H. Cox, N. L. Michael, L. Jagodzinski, C. Hawkes, D. Mayers, B. L. Gilliam, D. C. Birx, and M. L. Robb. 2001. Selective increases in HIV-specific neutralizing antibody and partial reconstitution of cellular immune responses during prolonged, successful drug therapy of HIV infection. AIDS Res. Hum. Retrovir. 17:1021-1034. - PubMed
-
- MedMira Laboratories, Inc. 2004. Reveal G2 Rapid HIV-1 Antibody Test. MedMira Laboratories, Inc., Halifax, Nova Scotia, Canada.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical