Pitfalls of intraoperative neuromonitoring for predicting postoperative recurrent laryngeal nerve function during thyroidectomy
- PMID: 16680596
- DOI: 10.1007/s00268-005-0355-8
Pitfalls of intraoperative neuromonitoring for predicting postoperative recurrent laryngeal nerve function during thyroidectomy
Abstract
Introduction: Intraoperative neuromonitoring has been widely adopted to facilitate the identification and preservation of recurrent laryngeal nerve (RLN) function during thyroid surgery. The present prospective study validated the ability of this technique to predict postoperative RLN outcomes in a single endocrine surgical unit.
Methods: Neuromonitoring was performed using Neurosign 100 with laryngeal surface electrodes in 171 patients with 271 nerves at risk during thyroidectomy. Vocal cord function was routinely documented perioperatively. Patients were also stratified to low risk (primary surgery for benign disease) and high risk (malignancy and recurrent disease) for subgroup analysis.
Results: Unilateral vocal cord palsy occurred in 15 patients (5.5%) postoperatively. The incidence of postoperative nerve palsy in the low risk and high risk groups was 4.4% and 7.8%, respectively. All but two patients had recovery of function within a median period of 4 months after the operation. The rates of transient and permanent RLN palsy based on nerves at risk were 4.8% (n = 13) and 0.7% (n = 2), respectively. There were 241 true-negative (positive signal and no cord palsy), 15 false-positive (negative signal but no cord palsy), 8 true-positive (negative signal and cord palsy), and 7 false-negative (positive signal but cord palsy) results, as correlated with the postoperative assessment. The sensitivity, specificity, and positive and negative predictive values were 53%, 94%, 35%, and 97%, respectively. For the high risk group, the sensitivity and positive predictive value increased to 86% and 60%, respectively.
Conclusions: There are pitfalls associated with the use of intraoperative neuromonitoring during thyroid surgery. Routine application is not recommended except for selected high risk patients.
Similar articles
-
[Value of the intraoperative neuromonitoring in surgery for thyroid cancer in identification and prognosis of function of the recurrent laryngeal nerves].Endokrynol Pol. 2006 Jul-Aug;57(4):343-6. Endokrynol Pol. 2006. PMID: 17006834 Polish.
-
Sensitivity and specificity of intraoperative recurrent laryngeal nerve stimulation test for predicting vocal cord palsy after thyroid surgery.World J Surg. 2006 Jul;30(7):1230-3. doi: 10.1007/s00268-005-0351-z. World J Surg. 2006. PMID: 16773263
-
[Intraoperative neuromonitoring during endocrine surgery].Ned Tijdschr Geneeskd. 2012;156(25):A4483. Ned Tijdschr Geneeskd. 2012. PMID: 22748368 Dutch.
-
Intraoperative monitoring of the recurrent laryngeal nerve in thyroid surgery.World J Surg. 2008 Jul;32(7):1358-66. doi: 10.1007/s00268-008-9483-2. World J Surg. 2008. PMID: 18305996 Review.
-
Comparison of Robotic Versus Conventional Open Thyroidectomy for Recurrent Laryngeal Nerve Safety: A Systematic Review and Meta-Analysis.Cureus. 2024 Feb 8;16(2):e53860. doi: 10.7759/cureus.53860. eCollection 2024 Feb. Cureus. 2024. PMID: 38465121 Free PMC article. Review.
Cited by
-
Electromyographic response persists after peripheral transection: endorsement of current concepts in recurrent laryngeal nerve monitoring in a porcine model.Langenbecks Arch Surg. 2010 Feb;395(2):121-5. doi: 10.1007/s00423-009-0570-0. Epub 2009 Nov 25. Langenbecks Arch Surg. 2010. PMID: 19937338
-
Results of intraoperative neuromonitoring in thyroid surgery and preoperative vocal cord paralysis.World J Surg. 2014 Mar;38(3):582-91. doi: 10.1007/s00268-013-2402-1. World J Surg. 2014. PMID: 24346632
-
Intraoperative nerve monitoring can reduce prevalence of recurrent laryngeal nerve injury in thyroid reoperations: results of a retrospective cohort study.World J Surg. 2014 Mar;38(3):599-606. doi: 10.1007/s00268-013-2260-x. World J Surg. 2014. PMID: 24081538 Free PMC article.
-
Intraoperative Nerve Monitoring in Thyroid Surgery: A Comprehensive Review of Technical Principles, Anesthetic Considerations, and Clinical Applications.J Clin Med. 2025 May 7;14(9):3259. doi: 10.3390/jcm14093259. J Clin Med. 2025. PMID: 40364290 Free PMC article. Review.
-
Anatomical variations of the recurrent laryngeal nerve in Chinese patients: a prospective study of 2,404 patients.Sci Rep. 2016 May 5;6:25475. doi: 10.1038/srep25475. Sci Rep. 2016. PMID: 27146369 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical