Laboratory productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions
- PMID: 16683868
- DOI: 10.5858/2006-130-596-LPATRO
Laboratory productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions
Abstract
Context: Automated laboratory hematology analyzers are capable of performing differential counts on peripheral blood smears with greater precision and more accurate detection of distributional and morphologic abnormalities than those performed by manual examinations of blood smears. Manual determinations of blood morphology and leukocyte differential counts are time-consuming, expensive, and may not always be necessary. The frequency with which hematology laboratory workers perform manual screens despite the availability of labor-saving features of automated analyzers is unknown.
Objective: To determine the normative rates with which manual peripheral blood smears were performed in clinical laboratories, to examine laboratory practices associated with higher or lower manual review rates, and to measure the effects of manual smear review on the efficiency of generating complete blood count (CBC) determinations.
Design: From each of 3 traditional shifts per day, participants were asked to select serially, 10 automated CBC specimens, and to indicate whether manual scans and/or reviews with complete differential counts were performed on blood smears prepared from those specimens. Sampling continued until a total of 60 peripheral smears were reviewed manually. For each specimen on which a manual review was performed, participants indicated the patient's age, hemoglobin value, white blood cell count, platelet count, and the primary reason why the manual review was performed. Participants also submitted data concerning their institutions' demographic profiles and their laboratories' staffing, work volume, and practices regarding CBC determinations. The rates of manual reviews and estimations of efficiency in performing CBC determinations were obtained from the data.
Setting: A total of 263 hospitals and independent laboratories, predominantly located in the United States, participating in the College of American Pathologists Q-Probes Program.
Results: There were 95,141 CBC determinations examined in this study; participants reviewed 15,423 (16.2%) peripheral blood smears manually. In the median institution (50th percentile), manual reviews of peripheral smears were performed on 26.7% of specimens. Manual differential count review rates were inversely associated with the magnitude of platelet counts that were required by laboratory policy to trigger smear reviews and with the efficiency of generating CBC reports. Lower manual differential count review rates were associated with laboratory policies that allowed manual reviews solely on the basis of abnormal automated red cell parameters and that precluded performing repeat manual reviews within designated time intervals. The manual scan rate elevated with increased number of hospital beds. In more than one third (35.7%) of the peripheral smears reviewed manually, participants claimed to have learned additional information beyond what was available on automated hematology analyzer printouts alone.
Conclusion: By adopting certain laboratory practices, it may be possible to reduce the rates of manual reviews of peripheral blood smears and increase the efficiency of generating CBC results.
Similar articles
-
Interinstitutional comparison of bedside blood glucose monitoring program characteristics, accuracy performance, and quality control documentation: a College of American Pathologists Q-Probes study of bedside blood glucose monitoring performed in 226 small hospitals.Arch Pathol Lab Med. 1998 Jun;122(6):495-502. Arch Pathol Lab Med. 1998. PMID: 9625416
-
Peripheral blood film review. The demise of the eyecount leukocyte differential.Clin Lab Med. 2002 Mar;22(1):279-97. doi: 10.1016/s0272-2712(03)00075-1. Clin Lab Med. 2002. PMID: 11933579 Review.
-
Morning rounds inpatient test availability: a College of American Pathologist Q-Probes study of 79860 morning complete blood cell count and electrolyte test results in 367 institutions.Arch Pathol Lab Med. 2000 Apr;124(4):499-503. doi: 10.5858/2000-124-0499-MRITA. Arch Pathol Lab Med. 2000. PMID: 10747303
-
[Current status and proposal of a guideline for manual slide review of automated complete blood cell count and white blood cell differential].Korean J Lab Med. 2010 Dec;30(6):559-66. doi: 10.3343/kjlm.2010.30.6.559. Korean J Lab Med. 2010. PMID: 21157139 Korean.
-
The international consensus group for hematology review: suggested criteria for action following automated CBC and WBC differential analysis.Lab Hematol. 2005;11(2):83-90. doi: 10.1532/LH96.05019. Lab Hematol. 2005. PMID: 16024331 Review.
Cited by
-
Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.Pract Lab Med. 2022 Apr 18;30:e00275. doi: 10.1016/j.plabm.2022.e00275. eCollection 2022 May. Pract Lab Med. 2022. PMID: 35601458 Free PMC article.
-
Comparison of haematological parameters determined by the Sysmex KX - 2IN automated haematology analyzer and the manual counts.BMC Clin Pathol. 2010 Apr 23;10:3. doi: 10.1186/1472-6890-10-3. BMC Clin Pathol. 2010. PMID: 20416068 Free PMC article.
-
Blood film in the era of streaming cells.Rev Bras Hematol Hemoter. 2017 Oct-Dec;39(4):295-296. doi: 10.1016/j.bjhh.2017.07.003. Epub 2017 Aug 23. Rev Bras Hematol Hemoter. 2017. PMID: 29150098 Free PMC article. No abstract available.
-
Hemogram study of an artificially feeding tree shrew (Tupaia belangeri chinensis).Exp Anim. 2020 Jan 29;69(1):80-91. doi: 10.1538/expanim.19-0079. Epub 2019 Sep 17. Exp Anim. 2020. PMID: 31527336 Free PMC article.
-
Are the review criteria for automated complete blood counts of the International Society of Laboratory Hematology suitable for all hematology laboratories?Rev Bras Hematol Hemoter. 2014 May-Jun;36(3):219-25. doi: 10.1016/j.bjhh.2014.03.011. Epub 2014 Apr 3. Rev Bras Hematol Hemoter. 2014. PMID: 25031063 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical