Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2006 Jul;52(7):1381-8.
doi: 10.1373/clinchem.2005.063735. Epub 2006 May 11.

Clinical-scale high-throughput analysis of urinary 8-oxo-7,8-dihydro-2'-deoxyguanosine by isotope-dilution liquid chromatography-tandem mass spectrometry with on-line solid-phase extraction

Affiliations

Clinical-scale high-throughput analysis of urinary 8-oxo-7,8-dihydro-2'-deoxyguanosine by isotope-dilution liquid chromatography-tandem mass spectrometry with on-line solid-phase extraction

Chiung-Wen Hu et al. Clin Chem. 2006 Jul.

Abstract

Background: Quantification of 8-oxo-7,8-dihydro-2'-deoxyguanosine (8-oxodGuo) in urine or blood is used to assess and monitor oxidative stress in patients. We describe the use of on-line solid-phase extraction (SPE) and isotope-dilution liquid chromatography-tandem mass spectrometry (LC-MS/MS) for automated measurement of urinary 8-oxodGuo.

Methods: Automated purification of urine was accomplished with a switching valve and an Inertsil ODS-3 column. After the addition of 15N5-labeled 8-oxodGuo as an internal standard, urine samples were analyzed within 10 min without sample purification. This method was applied to measure urinary 8-oxodGuo in a group of healthy persons (32 regular smokers and 35 nonsmokers). Urinary cotinine was also assayed by an isotope-dilution LC-MS/MS method.

Results: The lower limit of detection was 5.7 ng/L on column (2.0 fmol). Inter- and intraday imprecision (CV) was < 5.0%. Mean recovery of 8-oxodGuo in urine was 99%-102%. Mean (SD) urinary concentrations of 8-oxodGuo in smokers [7.26 (3.14) microg/g creatinine] were significantly higher than those in nonsmokers [4.69 (1.70) microg/g creatinine; P < 0.005]. Urinary concentrations of 8-oxodGuo were significantly correlated with concentrations of cotinine in smokers (P < 0.05).

Conclusions: This on-line SPE LC-MS/MS method is sufficiently sensitive, precise, and rapid to provide high-throughput direct analysis of urinary 8-oxodGuo without compromising quality and validation criteria. This method could be applicable for use in daily clinical practice for assessing oxidative stress in patients.

PubMed Disclaimer

Publication types