Comparative Study
. 2006 Jun;7(6):497-502.
doi: 10.1631/jzus.2006.B0497.

Evaluation of a kinetic uricase method for serum uric acid assay by predicting background absorbance of uricase reaction solution with an integrated method

Affiliations
Comparative Study

Evaluation of a kinetic uricase method for serum uric acid assay by predicting background absorbance of uricase reaction solution with an integrated method

Fei Liao et al. J Zhejiang Univ Sci B. 2006 Jun.

Abstract

A patented kinetic uricase method was evaluated for serum uric acid assay. Initial absorbance of the reaction mixture before uricase action (A(0)) was obtained by correcting the absorbance at 293 nm measured before the addition of uricase solution, and background absorbance (A(b)) was predicted by an integrated method. Uric acid concentration in reaction solution was calculated from A, the difference between A(0) and A(b), using the absorptivity preset for uric acid. This kinetic uricase method exhibited CV<4.3% and recovery of 100%. Lipids, bilirubin, hemoglobin, ascorbic acid, reduced glutathione and xanthine <0.32 mmol/L in serum had no significant effects. A linearly responded to 1.2 to 37.5 micromol/L uric acid in reaction solution containing 15 microl serum. The slope of linear response was consistent with the absorptivity preset for uric acid while the intercept was consistent with that for serum alone. Uric acid concentrations in clinic sera by different uricase methods positively correlated to each other. By Bland-Altman analysis, this kinetic uricase method accorded with that by quantifying the total change of UV absorbance on the completion of uricase reaction. These results demonstrated that this kinetic uricase method is reliable for serum uric acid assay with enhanced resistance to both xanthine and other common errors, wider range of linear response and much lower cost.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Bland-Altman analysis of the agreement between different uricase methods for the assay of uric acid in 189 clinic sera. (a) The indirect equilibrium method (C ie) vs the direct equilibrium method (C de). Bland-Altman analysis gave C ieC de=−0.002−0.073×((C ieC de)/2) (r>0.698, s=0.039) while the correlation equation between them was C ie=−0.001+0.927×C de (r>0.997, s=0.037); (b) This direct kinetic uricase method (C dk) vs the direct equilibrium method (C de). Bland-Altman analysis gave C dkC de=−0.019+0.010×((C de+C dk)/2) (r<0.137, s=0.037) while the correlation equation between them was C dk=−0.018+1.007×C de (r>0.998, s=0.037)
Fig. 1
Fig. 1
Bland-Altman analysis of the agreement between different uricase methods for the assay of uric acid in 189 clinic sera. (a) The indirect equilibrium method (C ie) vs the direct equilibrium method (C de). Bland-Altman analysis gave C ieC de=−0.002−0.073×((C ieC de)/2) (r>0.698, s=0.039) while the correlation equation between them was C ie=−0.001+0.927×C de (r>0.997, s=0.037); (b) This direct kinetic uricase method (C dk) vs the direct equilibrium method (C de). Bland-Altman analysis gave C dkC de=−0.019+0.010×((C de+C dk)/2) (r<0.137, s=0.037) while the correlation equation between them was C dk=−0.018+1.007×C de (r>0.998, s=0.037)

Similar articles

Cited by

References

    1. Aoki Y, Ihara H, Nakamura H, Aoki T, Yoshlda M. Effects of serum bilirubin on determination of uric acid by the uricase-peroxidase coupled reaction. Clin Chem. 1992;38(7):1350–1352. - PubMed
    1. Bartl K, Brandhuber M, Ziegenhorn J. Improved automated kinetic determination of uric acid in serum by use of uricase/catalase/aldehyde dehydrogenase. Clin Chem. 1979;25(4):619–621. - PubMed
    1. Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. The Lancet. 1995;346(8982):1085–1087. doi: 10.1016/S0140-6736(95)91748-9. - DOI - PubMed
    1. Duncan PH, Gochman N, Cooper T, Smith E, Bayse D. A candidate reference method for uric acid in serum. I. Optimization and evaluation. Clin Chem. 1982;28(2):284–290. - PubMed
    1. Hamilton SD, Pardue HL. Kinetic method having a linear range for substrate concentration that exceeds Michaelis-Menten constant. Clin Chem. 1982;28(12):2359–2365. - PubMed

Publication types