Organ movement reduction in PET/CT using dual-gated list-mode acquisition
- PMID: 16696375
- DOI: 10.1078/0939-3889-00296
Organ movement reduction in PET/CT using dual-gated list-mode acquisition
Abstract
Purpose: Imaging of moving organs using the PET leads to blurred images due to long acquisition times. Simultaneous cardiac and respiratory gating of list-mode PET/CT is evaluated with the aim to improve image quality and assess the organ movement.
Methods: We performed a N-13 ammonia PET/CT scan with a human volunteer, using the Siemens Biograph Sensation 16 scanner with list-mode acquisition. For ECG gating we used the scanner's integrated ECG device. Respiratory gating was done with the BioVet pneumatic sensor system.
Results: The sorting of the list-mode data post acquisition produced the desired matrix of eight cardiac times eight respiratory images. Organ movement could be measured in the series of gated PET images. The quantitation of tracer uptake in the myocardium showed artifacts due to the CT-based attenuation correction.
Conclusion: Double gating is feasible in human PET/CT scans using a list-mode-based scan protocol. The image quality can be enhanced using double gated list-mode acquisition in PET/CT Attenuation correction protocols in PET using a single not gated fast CT introduces artifacts in moving organs.
Similar articles
-
Detection of respiratory tumour motion using intrinsic list mode-driven gating in positron emission tomography.Eur J Nucl Med Mol Imaging. 2010 Dec;37(12):2315-27. doi: 10.1007/s00259-010-1533-y. Epub 2010 Jul 6. Eur J Nucl Med Mol Imaging. 2010. PMID: 20607534
-
Phased attenuation correction in respiration correlated computed tomography/positron emitted tomography.Med Phys. 2006 Jun;33(6):1840-7. doi: 10.1118/1.2198170. Med Phys. 2006. PMID: 16872091
-
Erroneous cardiac ECG-gated PET list-mode trigger events can be retrospectively identified and replaced by an offline reprocessing approach: first results in rodents.Phys Med Biol. 2013 Nov 21;58(22):7937-59. doi: 10.1088/0031-9155/58/22/7937. Epub 2013 Oct 29. Phys Med Biol. 2013. PMID: 24165267
-
New cardiac cameras: single-photon emission CT and PET.Semin Nucl Med. 2014 Jul;44(4):232-51. doi: 10.1053/j.semnuclmed.2014.04.003. Semin Nucl Med. 2014. PMID: 24948149 Review.
-
PET/CT Myocardial Perfusion Imaging Acquisition and Processing: Ten Tips and Tricks to Help You Succeed.Curr Cardiol Rep. 2021 Mar 11;23(5):39. doi: 10.1007/s11886-021-01476-5. Curr Cardiol Rep. 2021. PMID: 33694057 Review.
Cited by
-
Detection of respiratory tumour motion using intrinsic list mode-driven gating in positron emission tomography.Eur J Nucl Med Mol Imaging. 2010 Dec;37(12):2315-27. doi: 10.1007/s00259-010-1533-y. Epub 2010 Jul 6. Eur J Nucl Med Mol Imaging. 2010. PMID: 20607534
-
Respiratory-gated PET/CT versus delayed images for the quantitative evaluation of lower pulmonary and hepatic lesions.J Med Imaging Radiat Oncol. 2014;58(3):277-82. doi: 10.1111/1754-9485.12154. Epub 2014 Jan 20. J Med Imaging Radiat Oncol. 2014. PMID: 24438486 Free PMC article.
-
Estimation of optimal number of gates in dual gated 18F-FDG cardiac PET.Sci Rep. 2020 Nov 9;10(1):19362. doi: 10.1038/s41598-020-75613-5. Sci Rep. 2020. PMID: 33168859 Free PMC article.
-
Dual-gated cardiac PET-clinical feasibility study.Eur J Nucl Med Mol Imaging. 2010 Mar;37(3):505-16. doi: 10.1007/s00259-009-1252-4. Epub 2009 Sep 30. Eur J Nucl Med Mol Imaging. 2010. PMID: 19789872
-
Ultrafast Ultrasound Imaging for Super-Resolution Preclinical Cardiac PET.Mol Imaging Biol. 2020 Oct;22(5):1342-1352. doi: 10.1007/s11307-020-01512-w. Epub 2020 Jun 29. Mol Imaging Biol. 2020. PMID: 32602084 Free PMC article.
MeSH terms
LinkOut - more resources
Other Literature Sources