. 2006 May 24:6:7.
doi: 10.1186/1471-227X-6-7.

The insecure airway: a comparison of knots and commercial devices for securing endotracheal tubes

Affiliations

The insecure airway: a comparison of knots and commercial devices for securing endotracheal tubes

Paris B Lovett et al. BMC Emerg Med. .

Abstract

Background: Endotracheal Tubes (ETTs) are commonly secured using adhesive tape, cloth tape, or commercial devices. The objectives of the study were (1) To compare degrees of movement of ETTs secured with 6 different commercial devices and (2) To compare movement of ETTs secured with cloth tape tied with 3 different knots (hitches).

Methods: A 17 cm diameter PVC tube with 14 mm "mouth" hole in the side served as a mannequin. ETTs were subjected to repeated jerks, using a cable and pulley system.

Measurements: (1) Total movement of ETTs relative to "mouth" (measure used for devices) (2) Slippage of ETT through securing knot (measure used for knots).

Results: Among commercial devices, the Dale showed less movement than other devices, although some differences between devices did not reach significance. Among knots, Magnus and Clove Hitches produced less slippage than the Cow Hitch, but these differences did not reach statistical significance.

Conclusion: Among devices tested, the Dale was most secure. Within the scope offered by the small sample sizes, there were no statistically significant differences between the knots in this study.

PubMed Disclaimer

References

    1. Kapadia FN, Bajan KB, Raje KV. Airway accidents in intubated intensive care unit patients: an epidemiological study. Crit Care Med. 2000;28:659–64. doi: 10.1097/00003246-200003000-00010. - DOI - PubMed
    1. Mort TC. Unplanned tracheal extubation outside the operating room: a quality improvement audit of hemodynamic and tracheal airway complications associated with emergency tracheal reintubation. Anesth Analg. 1998;86:1171–6. doi: 10.1097/00000539-199806000-00006. - DOI - PubMed
    1. Boulain T. Unplanned extubations in the adult intensive care unit: a prospective multicenter study. Association des Reanimateurs du Centre-Ouest. Am J Respir Crit Care Med. 1998;157:1131–7. - PubMed
    1. Chevron V, Menard JF, Richard JC, Girault C, Leroy J, Bonmarchand G. Unplanned extubation: risk factors of development and predictive criteria for reintubation. Crit Care Med. 1998;26:1049–53. doi: 10.1097/00003246-199806000-00026. - DOI - PubMed
    1. Betbese AJ, Perez M, Bak E, Rialp G, Mancebo J. A prospective study of unplanned endotracheal extubation in intensive care unit patients. Crit Care Med. 1998;26:1180–6. doi: 10.1097/00003246-199807000-00016. - DOI - PubMed

LinkOut - more resources