Evaluation of isoflurane and sevoflurane vaporizers over a wide range of oxygen flow rates
- PMID: 16740083
- DOI: 10.2460/ajvr.67.6.936
Evaluation of isoflurane and sevoflurane vaporizers over a wide range of oxygen flow rates
Abstract
Objective: To examine the accuracy and precision of isoflurane and sevoflurane anesthetic vaporizers.
Sample population: 5 identical isoflurane and 5 identical sevoflurane vaporizers.
Procedures: Oxygen flow rates from 0.02 to 10 L/min were used with different vaporizer dial settings. Agent concentrations were measured at the common gas outlet by use of a refractometer. Accuracy was defined as the difference between measured agent concentrations, and dial settings were expressed as a percentage of the applied dial settings. Precision was defined as SD of the measured agent concentrations for each combination of dial setting and flow.
Results: Isoflurane values were generally greater than the dial settings. Accuracy of the isoflurane vaporizer was > 20% when 0.6% and 1% was dialed. Accuracy of the sevoflurane vaporizer was always within +/- 20% but decreased at 0.02 L/min flow and at combinations of high flow and high dial settings. Overall precision of the isoflurane vaporizer was better than that of the sevoflurane vaporizer.
Conclusions and clinical relevance: A possible explanation for the inaccuracy of the isoflurane vaporizer may be that it was manufactured to be accurate with air but not oxygen, which must be accounted for when using the vaporizer with oxygen, especially with nonrebreathing systems. The sevoflurane vaporizer may not deliver accurate agent concentrations at high flow and high dial settings. Both vaporizers are suitable for clinical use with a wide range of oxygen flow rates if these precautions are properly addressed.
Similar articles
-
Accuracy of isoflurane, halothane, and sevoflurane vaporizers during high oxygen flow and at maximum vaporizer dial setting.Am J Vet Res. 2011 Jun;72(6):751-6. doi: 10.2460/ajvr.72.6.751. Am J Vet Res. 2011. PMID: 21627520
-
Accuracy of isoflurane delivery by halothane-specific vaporizers.Am J Vet Res. 1983 Jun;44(6):1072-8. Am J Vet Res. 1983. PMID: 6870010
-
Evaluation of an isoflurane vaporizer: the Cyprane Fortec.Anesth Analg. 1982 May;61(5):457-64. Anesth Analg. 1982. PMID: 7039418
-
[Anesthesia equipment: fresh gas delivery systems. I. Mecanical systems with rotameters and calibrated vaporizers].Ann Fr Anesth Reanim. 1999 Nov;18(9):956-75. doi: 10.1016/s0750-7658(00)87945-9. Ann Fr Anesth Reanim. 1999. PMID: 10615544 Review. French.
-
Vaporizer position.Vet Clin North Am Small Anim Pract. 1992 Mar;22(2):389-91. doi: 10.1016/s0195-5616(92)50645-3. Vet Clin North Am Small Anim Pract. 1992. PMID: 1585583 Review.
Cited by
-
Use of a Low-flow Digital Anesthesia System for Mice and Rats.J Vis Exp. 2016 Sep 7;(115):54436. doi: 10.3791/54436. J Vis Exp. 2016. PMID: 27684327 Free PMC article.
-
Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents.J Vis Exp. 2020 Jul 9;(161):10.3791/61311. doi: 10.3791/61311. J Vis Exp. 2020. PMID: 32716388 Free PMC article.
-
Do the Manual or Computer-Controlled Flowmeters Generate Similar Isoflurane Concentrations in Tafonius?Front Vet Sci. 2019 May 29;6:160. doi: 10.3389/fvets.2019.00160. eCollection 2019. Front Vet Sci. 2019. PMID: 31192237 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical