. 2006 Jul;17(7):977-982.
doi: 10.1016/j.jasms.2006.03.006. Epub 2006 Jun 5.

Dynamic range of mass accuracy in LTQ Orbitrap hybrid mass spectrometer

Affiliations

Dynamic range of mass accuracy in LTQ Orbitrap hybrid mass spectrometer

Alexander Makarov et al. J Am Soc Mass Spectrom. 2006 Jul.

Erratum in

Abstract

Using a novel orbitrap mass spectrometer, the authors investigate the dynamic range over which accurate masses can be determined (extent of mass accuracy) for short duration experiments typical for LC/MS. A linear ion trap is used to selectively fill an intermediate ion storage device (C-trap) with ions of interest, following which the ensemble of ions is injected into an orbitrap mass analyzer and analyzed using image current detection and fast Fourier transformation. Using this technique, it is possible to generate ion populations with intraspectrum intensity ranges up to 10(4). All measurements (including ion accumulation and image current detection) were performed in less than 1 s at a resolving power of 30,000. It was shown that 5-ppm mass accuracy of the orbitrap mass analyzer is reached with >95% probability at a dynamic range of more than 5000, which is at least an order of magnitude higher than typical values for time-of-flight instruments. Due to the high resolving power of the orbitrap, accurate mass of an ion could be determined when the signal was reliably distinguished from noise (S/Np-p)>2...3).

PubMed Disclaimer

References

    1. J Proteome Res. 2004 May-Jun;3(3):621-6 - PubMed
    1. Rapid Commun Mass Spectrom. 2004;18(4):511-7 - PubMed
    1. Anal Chem. 2000 Mar 15;72(6):1156-62 - PubMed
    1. J Mass Spectrom. 2003 Oct;38(10):1043-53 - PubMed
    1. Anal Chem. 1984 Dec;56(14 ):2744-8 - PubMed

LinkOut - more resources