Minimization of cochlear implant stimulus artifact in cortical auditory evoked potentials
- PMID: 16807102
- DOI: 10.1016/j.clinph.2006.04.018
Minimization of cochlear implant stimulus artifact in cortical auditory evoked potentials
Abstract
Objective: To compare two methods of minimizing cochlear implant artifact in cortical auditory evoked potential (CAEP) recordings.
Methods: Two experiments were conducted. In the first, we assessed the use of independent component analysis (ICA) as a pre-processing filter. In the second, we explored the use of an optimized differential reference (ODR) for minimizing artifacts.
Results: Both ICA and the ODR can minimize the artifact and allow measurement of CAEP responses.
Conclusions: When using a large number of recording electrodes ICA can be used to minimize the implant artifact. When using a single electrode montage an optimized differential reference is adequate to minimize the artifact.
Significance: The use of an optimized differential reference could allow cortical evoked potentials to be used in routine clinical assessment of auditory pathway development in children and adults fit with cochlear implants.
Similar articles
-
Minimization of cochlear implant artifact in cortical auditory evoked potentials in children.Int J Pediatr Otorhinolaryngol. 2012 Nov;76(11):1627-32. doi: 10.1016/j.ijporl.2012.07.034. Epub 2012 Aug 19. Int J Pediatr Otorhinolaryngol. 2012. PMID: 22910837
-
Characterizing responses from auditory cortex in young people with several years of cochlear implant experience.Clin Neurophysiol. 2008 Oct;119(10):2347-62. doi: 10.1016/j.clinph.2008.06.013. Epub 2008 Aug 26. Clin Neurophysiol. 2008. PMID: 18752993
-
A method for removing cochlear implant artifact.Hear Res. 2010 Jan;259(1-2):95-106. doi: 10.1016/j.heares.2009.10.012. Epub 2009 Oct 28. Hear Res. 2010. PMID: 19878712
-
Maturation of CAEP in infants and children: a review.Hear Res. 2006 Feb;212(1-2):212-23. doi: 10.1016/j.heares.2005.11.008. Epub 2006 Feb 9. Hear Res. 2006. PMID: 16480841 Review.
-
The role of sound in adult and developmental auditory cortical plasticity.Ear Hear. 2008 Dec;29(6):819-29. doi: 10.1097/AUD.0b013e3181853030. Ear Hear. 2008. PMID: 18941413 Review.
Cited by
-
Effects of Stimulus Duration on Event-Related Potentials Recorded From Cochlear-Implant Users.Ear Hear. 2017 Nov/Dec;38(6):e389-e393. doi: 10.1097/AUD.0000000000000444. Ear Hear. 2017. PMID: 28475545 Free PMC article.
-
Validation of direct recording of electrically evoked cortical auditory evoked potentials through a cochlear implant system.Sci Rep. 2024 Nov 17;14(1):28366. doi: 10.1038/s41598-024-79528-3. Sci Rep. 2024. PMID: 39551893 Free PMC article.
-
Cortical reorganization in children with cochlear implants.Brain Res. 2008 Nov 6;1239:56-65. doi: 10.1016/j.brainres.2008.08.026. Epub 2008 Aug 18. Brain Res. 2008. PMID: 18775684 Free PMC article.
-
The Effects of Cortical Reorganization and Applications of Functional Near-Infrared Spectroscopy in Deaf People and Cochlear Implant Users.Brain Sci. 2022 Aug 28;12(9):1150. doi: 10.3390/brainsci12091150. Brain Sci. 2022. PMID: 36138885 Free PMC article. Review.
-
Interaural speech asymmetry predicts bilateral speech intelligibility but not listening effort in adults with bilateral cochlear implants.Front Neurosci. 2022 Dec 7;16:1038856. doi: 10.3389/fnins.2022.1038856. eCollection 2022. Front Neurosci. 2022. PMID: 36570844 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous