The purpose of this study was to provide a quantitative assessment of galvanic corrosion behavior of orthodontic archwire alloys coupled to orthodontic bracket alloys in 0.9% NaCl solution and to study the effect of surface area ratios. Two common bracket alloys, stainless steels and titanium, and four common wire alloys, nickel-titanium (NiTi) alloy, beta-titanium (beta-Ti) alloy, stainless steel, and cobalt-chromium-nickel alloy, were used. Three different area ratios, 1:1, 1:2.35, and 1:3.64, were used; two of them assumed that the multibracket appliances consists of 14 brackets and 0.016 inch of round archwire or 0.016 x 0.022 inch of rectangular archwire. The galvanic current was measured for 3 successive days using zero-impedance ammeter. When the NiTi alloy was coupled with Ti (1:1, 1:2.35, and 1:3.64 of the surface area ratio) or beta-Ti alloy was coupled with Ti (1:2.35 and 1:3.64 of the surface area ratio), Ti initially was the anode and corroded. However, the polarity reversed in 1 hour, resulting in corrosion of the NiTi or beta-Ti. The NiTi alloy coupled with SUS 304 or Ti exhibited a relatively large galvanic current density even after 72 hours. It is suggested that coupling SUS 304-NiTi and Ti-NiTi may remarkably accelerate the corrosion of NiTi alloy, which serves as the anode. The different anode-cathode area ratios used in this study had little effect on galvanic corrosion behavior.
Similar articles
-
An investigation into the effects of polishing on surface hardness and corrosion of orthodontic archwires.Angle Orthod. 1999 Oct;69(5):433-40. doi: 10.1043/0003-3219(1999)069<0433:AIITEO>2.3.CO;2. Angle Orthod. 1999. PMID: 10515141
-
In vitro surface corrosion of stainless steel and NiTi orthodontic appliances.Aust Orthod J. 2003 Apr;19(1):13-8. Aust Orthod J. 2003. PMID: 12790351
-
Comparison of galvanic corrosion potential of metal injection molded brackets to that of conventional metal brackets with nickel-titanium and copper nickel-titanium archwire combinations.J Contemp Dent Pract. 2013 May 1;14(3):488-95. doi: 10.5005/jp-journals-10024-1350. J Contemp Dent Pract. 2013. PMID: 24171995
-
In vivo aging of orthodontic alloys: implications for corrosion potential, nickel release, and biocompatibility.Angle Orthod. 2002 Jun;72(3):222-37. doi: 10.1043/0003-3219(2002)072<0222:IVAOOA>2.0.CO;2. Angle Orthod. 2002. PMID: 12071606 Review.
-
Coefficients of friction for arch wires in stainless steel and polycrystalline alumina bracket slots. I. The dry state.Am J Orthod Dentofacial Orthop. 1990 Oct;98(4):300-12. doi: 10.1016/S0889-5406(05)81487-8. Am J Orthod Dentofacial Orthop. 1990. PMID: 2220691 Review.
Cited by
-
Galvanic coupling of steel and gold alloy lingual brackets with orthodontic wires: Is corrosion a concern?Angle Orthod. 2018 Jul;88(4):450-457. doi: 10.2319/092917-655.1. Epub 2018 Mar 6. Angle Orthod. 2018. PMID: 29509025 Free PMC article.
-
Virulence modulation of Streptococcus mutans biofilms by metal ions released from orthodontic appliances.Angle Orthod. 2013 Nov;83(6):987-93. doi: 10.2319/112712-904.1. Epub 2013 Apr 15. Angle Orthod. 2013. PMID: 23586445 Free PMC article.
-
Release of nickel and chromium ions from orthodontic wires following the use of teeth whitening mouthwashes.Prog Orthod. 2018 Feb 5;19(1):4. doi: 10.1186/s40510-018-0203-7. Prog Orthod. 2018. PMID: 29399703 Free PMC article.
-
Minimization of Adverse Effects Associated with Dental Alloys.Materials (Basel). 2022 Oct 25;15(21):7476. doi: 10.3390/ma15217476. Materials (Basel). 2022. PMID: 36363067 Free PMC article. Review.
-
In vitro evaluation of physical vapor deposition coated beta titanium orthodontic archwires.Angle Orthod. 2012 Jan;82(1):22-9. doi: 10.2319/040811-251.1. Epub 2011 Jul 12. Angle Orthod. 2012. PMID: 21749248 Free PMC article.