Monitoring buried free flaps: limitations of the implantable Doppler and use of color duplex sonography as a confirmatory test
- PMID: 16816680
- DOI: 10.1097/01.prs.0000221113.78244.8c
Monitoring buried free flaps: limitations of the implantable Doppler and use of color duplex sonography as a confirmatory test
Abstract
Background: Postoperative monitoring of free flaps is important to minimize the risk of flap failure, but monitoring buried free flaps is difficult because the standard methods of clinical examination and surface Doppler monitoring are not possible. Buried free flaps are often monitored using an implantable 20-MHz ultrasonic Doppler probe.
Methods: The authors conducted a retrospective clinical study of buried free flaps to assess the reliability of the implantable Doppler probe in postoperative monitoring of free flaps.
Results: During the 38-month study period, 956 free flap operations were performed at the authors' institution. Twenty (2.1 percent) of these cases involved completely buried free flaps in which an implantable Doppler probe was used for flap monitoring. Implantable Doppler probe monitoring had a 100 percent sensitivity rate in detecting loss of flap perfusion, making it a good screening test for free flap viability. However, it suffered from a high false-positive rate of 88 percent, which resulted in a high proportion of subsequent negative surgical explorations. In one case, color duplex sonography, a rapid and noninvasive test, revealed that the loss of signal from the implantable Doppler probe was a false-positive result.
Conclusions: The implantable Doppler probe is a sensitive method for postoperative monitoring of free flaps but is prone to false-positive signals. The use of color duplex sonography to confirm implantable Doppler probe findings may avert unnecessary surgical exploration, thereby improving postoperative monitoring of free flaps.
Similar articles
-
A chimaeric-pattern flap design for implantable Doppler surrogate monitoring: a novel placement technique.J Plast Reconstr Aesthet Surg. 2014 Feb;67(2):190-7. doi: 10.1016/j.bjps.2013.10.045. Epub 2013 Nov 13. J Plast Reconstr Aesthet Surg. 2014. PMID: 24365566
-
Color duplex sonography for the monitoring of vascularized free bone flaps.Otolaryngol Head Neck Surg. 2003 Jul;129(1):71-6. doi: 10.1016/S0194-59980300486-8. Otolaryngol Head Neck Surg. 2003. PMID: 12869920
-
Efficacy of conventional monitoring techniques in free tissue transfer: an 11-year experience in 750 consecutive cases.Plast Reconstr Surg. 1999 Jul;104(1):97-101. Plast Reconstr Surg. 1999. PMID: 10597680
-
Advancements in free flap monitoring in the last decade: a critical review.Plast Reconstr Surg. 2010 Jan;125(1):177-185. doi: 10.1097/PRS.0b013e3181c49580. Plast Reconstr Surg. 2010. PMID: 20048610 Review.
-
Modern postoperative monitoring of free flaps.Curr Opin Otolaryngol Head Neck Surg. 2018 Aug;26(4):248-253. doi: 10.1097/MOO.0000000000000467. Curr Opin Otolaryngol Head Neck Surg. 2018. PMID: 29750670 Review.
Cited by
-
Implantable Doppler Probes for Postoperatively Monitoring Free Flaps: Efficacy. A Systematic Review and Meta-analysis.Plast Reconstr Surg Glob Open. 2016 Nov 28;4(11):e1099. doi: 10.1097/GOX.0000000000001099. eCollection 2016 Nov. Plast Reconstr Surg Glob Open. 2016. PMID: 27975015 Free PMC article.
-
Postoperative Free-Flap Monitoring Techniques.Semin Plast Surg. 2019 Feb;33(1):13-16. doi: 10.1055/s-0039-1677880. Epub 2019 Mar 8. Semin Plast Surg. 2019. PMID: 30863207 Free PMC article. Review.
-
Venous Occlusion Detected by Caregiver with Implantable Doppler in a Buried Free Flap.Arch Craniofac Surg. 2014 Dec;15(3):121-124. doi: 10.7181/acfs.2014.15.3.121. Epub 2014 Dec 23. Arch Craniofac Surg. 2014. PMID: 28913204 Free PMC article.
-
A System for Simple Real-Time Anastomotic Failure Detection and Wireless Blood Flow Monitoring in the Lower Limbs.IEEE J Transl Eng Health Med. 2016 Aug 25;4:4100114. doi: 10.1109/JTEHM.2016.2588504. eCollection 2016. IEEE J Transl Eng Health Med. 2016. PMID: 27730016 Free PMC article.
-
Efficacy of indocyanine green fluorescence-based near-infrared angiography in assessing intraoperative wound perfusion for bone and soft-tissue surgery.Bone Jt Open. 2025 Jul 11;6(7):796-806. doi: 10.1302/2633-1462.67.BJO-2024-0248.R1. Bone Jt Open. 2025. PMID: 40641409 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources