The accuracy of 1- and 3-mm slices in coronary calcium scoring using multi-slice CT in vitro and in vivo
- PMID: 16819606
- DOI: 10.1007/s00330-006-0332-0
The accuracy of 1- and 3-mm slices in coronary calcium scoring using multi-slice CT in vitro and in vivo
Abstract
The accuracy of coronary calcium scoring using 16-row MSCT comparing 1- and 3-mm slices was assessed. A thorax phantom with calcium cylinder inserts was scanned applying a non-enhanced retrospectively ECG-gated examination protocol: collimation 12 x 0.75 mm; 120 kV; 133 mAs(eff). Thirty-eight patients were examined using the same scan protocol. Image reconstruction was performed with an effective slice thickness of 3 and 1 mm. The volume score, calcium mass and Agatston score were determined. Image noise was measured in both studies. The volume score and calcium mass varied less than the Agatston score. The overall measured calcium mass compared to the actual calcium mass revealed a relative difference of +2.0% for 1-mm slices and -1.2% for 3-mm slices. Due to increased image noise in thinner slices in the patient study (26.1 HU), overall calcium scoring with a scoring threshold of 130 HU was not feasible. Interlesion comparison showed significantly higher scoring results for thinner slices (all P<0.001). A similar accuracy comparing calcium scoring results of 1- and 3-mm slices was shown in the phantom study; therefore, the potentially necessary increase of the patient's dose in order to achieve assessable 1-mm slices with an acceptable image-to-noise-ratio appears not to be justified.
Similar articles
-
Effect of varying slice thickness on coronary calcium scoring with multislice computed tomography in vitro and in vivo.Invest Radiol. 2005 Nov;40(11):695-9. doi: 10.1097/01.rli.0000179523.07907.a6. Invest Radiol. 2005. PMID: 16230901
-
Very low-dose coronary artery calcium scanning with high-pitch spiral acquisition mode: comparison between 120-kV and 100-kV tube voltage protocols.J Cardiovasc Comput Tomogr. 2013 Jan-Feb;7(1):32-8. doi: 10.1016/j.jcct.2012.11.004. Epub 2012 Dec 1. J Cardiovasc Comput Tomogr. 2013. PMID: 23333186
-
Calcium score of small coronary calcifications on multidetector computed tomography: results from a static phantom study.Eur J Radiol. 2013 Feb;82(2):e58-63. doi: 10.1016/j.ejrad.2012.09.018. Epub 2012 Oct 23. Eur J Radiol. 2013. PMID: 23092538
-
Assessment of calcium scoring performance in cardiac computed tomography.Eur Radiol. 2003 Mar;13(3):484-97. doi: 10.1007/s00330-002-1746-y. Epub 2002 Dec 4. Eur Radiol. 2003. PMID: 12594550
-
[Detection of coronary calcinosis with multislice spiral computerized tomography: an alternative to electron beam tomography].Z Kardiol. 2002 Aug;91(8):642-9. doi: 10.1007/s00392-002-0838-7. Z Kardiol. 2002. PMID: 12426828 Clinical Trial. German.
Cited by
-
Image quality of multiplanar reconstruction of pulmonary CT scans using adaptive statistical iterative reconstruction.Br J Radiol. 2011 Apr;84(1000):335-41. doi: 10.1259/bjr/57998586. Epub 2010 Nov 16. Br J Radiol. 2011. PMID: 21081572 Free PMC article.
-
CT attenuation values of blood and myocardium: rationale for accurate coronary artery calcifications detection with multi-detector CT.PLoS One. 2015 Apr 14;10(4):e0124175. doi: 10.1371/journal.pone.0124175. eCollection 2015. PLoS One. 2015. PMID: 25875629 Free PMC article.
-
Validation of a Length-Adjusted Abdominal Arterial Calcium Score Method for Contrast-Enhanced CT Scans.Diagnostics (Basel). 2023 Jun 1;13(11):1934. doi: 10.3390/diagnostics13111934. Diagnostics (Basel). 2023. PMID: 37296786 Free PMC article.
-
Coronary calcification and bone microarchitecture by high-resolution peripheral quantitative computed tomography from the São Paulo Ageing and Health (SPAH) Study.Sci Rep. 2022 Mar 28;12(1):5282. doi: 10.1038/s41598-022-08839-0. Sci Rep. 2022. PMID: 35347151 Free PMC article.
-
Measurement of coronary calcium scores by electron beam computed tomography or exercise testing as initial diagnostic tool in low-risk patients with suspected coronary artery disease.Eur Radiol. 2008 Feb;18(2):244-52. doi: 10.1007/s00330-007-0755-2. Epub 2007 Sep 28. Eur Radiol. 2008. PMID: 17901959 Free PMC article. Clinical Trial.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical