. 2006 Aug;385(8):1520-5.
doi: 10.1007/s00216-006-0554-7. Epub 2006 Jul 4.

Sensitive determination of thiamethoxam, imidacloprid and acetamiprid in environmental water samples with solid-phase extraction packed with multiwalled carbon nanotubes prior to high-performance liquid chromatography

Affiliations

Sensitive determination of thiamethoxam, imidacloprid and acetamiprid in environmental water samples with solid-phase extraction packed with multiwalled carbon nanotubes prior to high-performance liquid chromatography

Qingxiang Zhou et al. Anal Bioanal Chem. 2006 Aug.

Abstract

This paper describes a novel method for sensitive determination of thiamethoxam, imidacloprid and acetamiprid based on solid-phase extraction with multiwalled carbon nanotubes as the packed materials. Factors that maybe influence the enrichment efficiency, such as sample flow rate, sample pH, and sample volume, were investigated in detail. Under the optimized conditions, the detection limits of thiamethoxam, imidacloprid and acetamiprid were 6.1, 5.4 and 6.7 ng L(-1), respectively. The experimental results indicated that there was good linearity (R2>0.9993) over the range of 0.08 approximately 100 ng mL(-1) and good reproducibility with the relative standard deviations over the range of 0.7 approximately 1.1% (n=6). The proposed method has been applied to the analysis of real-world water samples, and satisfactory achievements were obtained. The average spiked recoveries were in the range of 87.5 approximately 109.8%. All the results indicated that the proposed method could be used for the simultaneous determination of the three pesticides in environmental water samples at trace levels.