Determination of LiF:Mg,Ti and LiF:Mg,Cu,P TL efficiency for X-rays and their application to Monte Carlo simulations of dosemeter response
- PMID: 16822775
- DOI: 10.1093/rpd/ncj001
Determination of LiF:Mg,Ti and LiF:Mg,Cu,P TL efficiency for X-rays and their application to Monte Carlo simulations of dosemeter response
Abstract
The simulation of response of a new passive area dosemeter for measuring ambient dose equivalent H*(10) for photons has been performed using the Monte Carlo code MCNP and experimentally determined responses of LiF:Mg,Ti and LiF:Mg,Cu,P thermoluminescent (TL) detectors for hard-filtered X-ray spectra from 20 to 300 keV and for 137Cs and 60Co gamma radiation. Relative TL efficiency for both types of detectors, determined in experiments with bare detectors and similar Monte Carlo simulations, compared favourably with prediction of microdosimetric models for proposed microdosimetric target sizes in the range of 20-40 nm. The concluding verification experiment showed small deviations between measured and simulated dosemeter energy response values in the range of a few percent.
Similar articles
-
Energy response of a two-dimensional sheet-type LiF:Mg,Cu,P TL dosemeter to photons.Radiat Prot Dosimetry. 2006;120(1-4):125-8. doi: 10.1093/rpd/nci527. Epub 2006 Apr 13. Radiat Prot Dosimetry. 2006. PMID: 16614093
-
The application of LiF:Mg,Cu,P to large scale personnel dosimetry: current status and future directions.Radiat Prot Dosimetry. 2006;119(1-4):248-54. doi: 10.1093/rpd/nci692. Epub 2006 Jul 11. Radiat Prot Dosimetry. 2006. PMID: 16835277 Review.
-
Comparison of two extremity dosemeters based on LiF:Mg,Cu,P thin detectors for mixed beta-gamma fields.Radiat Prot Dosimetry. 2006;120(1-4):316-20. doi: 10.1093/rpd/nci586. Epub 2006 Apr 27. Radiat Prot Dosimetry. 2006. PMID: 16644980
-
Comparison of the energy-response factor of LiF and Al2O3 in radiotherapy beams.Radiat Prot Dosimetry. 2006;119(1-4):497-9. doi: 10.1093/rpd/nci676. Epub 2006 May 30. Radiat Prot Dosimetry. 2006. PMID: 16735558
-
The effects of ionisation density on the thermoluminescence response (efficiency) of LiF:Mg,Ti and LiF:Mg,Cu,P.Radiat Prot Dosimetry. 2004;109(4):331-48. doi: 10.1093/rpd/nch310. Radiat Prot Dosimetry. 2004. PMID: 15273352 Review.
Cited by
-
Radiation exposure to the orthopaedic surgeon during periacetabular osteotomy.Int Orthop. 2009 Dec;33(6):1747-51. doi: 10.1007/s00264-008-0681-1. Epub 2008 Oct 29. Int Orthop. 2009. PMID: 18958467 Free PMC article.
-
Comparison of TLD calibration methods for 192Ir dosimetry.J Appl Clin Med Phys. 2013 Jan 7;14(1):4037. doi: 10.1120/jacmp.v14i1.4037. J Appl Clin Med Phys. 2013. PMID: 23318392 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources