Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose-response using different monomers
- PMID: 16825731
- DOI: 10.1088/0031-9155/51/14/001
Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose-response using different monomers
Abstract
In this work, three new polymer gel dosimeter recipes were investigated that may be more suitable for widespread applications than polyacrylamide gel dosimeters, since the extremely toxic acrylamide has been replaced with the less harmful monomers N-isopropylacrylamide (NIPAM), diacetone acrylamide and N-vinylformamide. The new gel dosimeters studied contained gelatin (5 wt%), monomer (3 wt%), N,N'-methylene-bis-acrylamide crosslinker (3 wt%) and tetrakis (hydroxymethyl) phosphonium chloride antioxidant (10 mM). The NMR response (R2) of the dosimeters was analysed for conditions of varying dose, dose rate, time post-irradiation, and temperature during irradiation and scanning. It was shown that the dose-response behaviour of the NIPAM/Bis gel dosimeter is comparable to that of normoxic polyacrylamide gel (PAGAT) in terms of high dose-sensitivity and low dependence on dose rate and irradiation temperature, within the ranges considered. The dose-response (R2) of NIPAM/Bis appears to be linear over a greater dose range than the PAGAT gel dosimeter. The effects of time post-irradiation (temporal instability) and temperature during NMR scanning on the R2 response were more significant for NIPAM/Bis dosimeters. Diacetone acrylamide and N-vinylformamide gel dosimeters possessed considerably lower dose-sensitivities. The optical dose-response, measured in terms of the attenuation coefficient for each polymer gel dosimeter, showed potential for the use of optical imaging techniques in future studies.
Similar articles
-
Characteristics of a novel polymer gel dosimeter formula for MRI scanning: Dosimetry, toxicity and temporal stability of response.Phys Med. 2016 Sep;32(9):1156-61. doi: 10.1016/j.ejmp.2016.08.008. Epub 2016 Aug 16. Phys Med. 2016. PMID: 27542576
-
An experimental study of the dose response of polymer gel dosimeters imaged with x-ray computed tomography.Phys Med Biol. 2001 Nov;46(11):2939-51. doi: 10.1088/0031-9155/46/11/312. Phys Med Biol. 2001. PMID: 11720356
-
Investigation of the PAGAT polymer gel dosimeter using magnetic resonance imaging.Phys Med Biol. 2005 Aug 21;50(16):3875-88. doi: 10.1088/0031-9155/50/16/015. Epub 2005 Aug 2. Phys Med Biol. 2005. PMID: 16077233
-
Radiation dosimetry using polymer gels: methods and applications.Br J Radiol. 2000 Sep;73(873):919-29. doi: 10.1259/bjr.73.873.11064643. Br J Radiol. 2000. PMID: 11064643 Review.
-
A comprehensive review of the gel dosimeters response at different energy levels and dose rates.Appl Radiat Isot. 2025 Jul;221:111800. doi: 10.1016/j.apradiso.2025.111800. Epub 2025 Mar 21. Appl Radiat Isot. 2025. PMID: 40188777 Review.
Cited by
-
Influence of magnesium chloride on the dose-response of polyacrylamide-type gel dosimeters.Radiol Phys Technol. 2018 Dec;11(4):375-381. doi: 10.1007/s12194-018-0473-2. Epub 2018 Sep 4. Radiol Phys Technol. 2018. PMID: 30182145
-
Improvement of sensitivity of X-ray CT reading method for polymer gel in radiation therapy.Rep Pract Oncol Radiother. 2020 Jan-Feb;25(1):100-103. doi: 10.1016/j.rpor.2019.12.017. Epub 2019 Dec 19. Rep Pract Oncol Radiother. 2020. PMID: 31908601 Free PMC article.
-
Design and construction of an optical computed tomography scanner for polymer gel dosimetry application.J Med Signals Sens. 2014 Apr;4(2):130-8. J Med Signals Sens. 2014. PMID: 24761377 Free PMC article.
-
Effect of substrate storage conditions on the stability of "Smart" films used for mammalian cell applications.Appl Surf Sci. 2017 Jan 15;392:950-959. doi: 10.1016/j.apsusc.2016.07.004. Epub 2016 Sep 1. Appl Surf Sci. 2017. PMID: 29081564 Free PMC article.
-
Basic Properties of a New Polymer Gel for 3D-Dosimetry at High Dose-Rates Typical for FFF Irradiation Based on Dithiothreitol and Methacrylic Acid (MAGADIT): Sensitivity, Range, Reproducibility, Accuracy, Dose Rate Effect and Impact of Oxygen Scavenger.Polymers (Basel). 2019 Oct 19;11(10):1717. doi: 10.3390/polym11101717. Polymers (Basel). 2019. PMID: 31635117 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous