The measurement of accurate fetal hemoglobin and related oxygen saturation by the hemoximeter
- PMID: 16844104
- PMCID: PMC1950590
- DOI: 10.1016/j.cca.2006.05.042
The measurement of accurate fetal hemoglobin and related oxygen saturation by the hemoximeter
Abstract
Background: The purposes of this study were to examine the accuracy of fetal hemoglobin (HbF) as quickly measured by the hemoximeter, verified by the high-performance liquid chromatography method, and to examine related oxygen saturation (SO(2)) measurements in neonates.
Methods: Thirty-nine neonates with gestational ages ranging from 25 to 38 weeks were investigated (n=280 blood samples). Twenty younger premature neonates had blood transfusions (n=188 blood samples, 72 before and 116 after transfusions), and 19 older neonates did not.
Results: The bias of the hemoximeter was 23% (+/-9.1) against the HPLC; 25% (+/-7.9) before, and 19% (+/-8.6) after blood transfusions (all P<0.001), for HbF measurements. A regression line (HbFt by the HPLC=8.46+0.7 x HbF by the hemoximeter) has been provided for the prediction. Oxyhemoglobin dissociation curves with the status of (before and after) blood transfusions were presented. In relation to oxygen tension values of 50-75 mm Hg, in addition to the right-shifted oxyhemoglobin dissociation curves, pulse oximeter ranged from 95 to 98% before the transfusions, but decreased to 94 to 96% after the blood transfusions.
Conclusions: Accurate HbF and related oxygen saturation measurements need to be determined, especially for premature neonates, to minimize the risk of oxygen toxicity.
Figures


Similar articles
-
Effects of fetal hemoglobin on accurate measurements of oxygen saturation in neonates.J Perinat Neonatal Nurs. 2005 Oct-Dec;19(4):348-61. doi: 10.1097/00005237-200510000-00010. J Perinat Neonatal Nurs. 2005. PMID: 16292136 Clinical Trial.
-
Accurate measurements of fetal hemoglobin for neonates with different gestational ages.Hemoglobin. 2006;30(4):419-35. doi: 10.1080/03630260600867883. Hemoglobin. 2006. PMID: 16987797 Clinical Trial.
-
From oxygen content to pulse oximetry: completing the picture in the newborn.Acta Anaesthesiol Scand Suppl. 1995;107:95-100. doi: 10.1111/j.1399-6576.1995.tb04341.x. Acta Anaesthesiol Scand Suppl. 1995. PMID: 8599308 Review.
-
Effect of fetal haemoglobin on the accuracy of pulse oximetry in preterm infants.J Paediatr Child Health. 1992 Feb;28(1):43-6. doi: 10.1111/j.1440-1754.1992.tb02615.x. J Paediatr Child Health. 1992. PMID: 1372809
-
Oxygen saturation in pulse oximetry in hemoglobin anomalies.Klin Padiatr. 2012 Jul;224(4):259-65. doi: 10.1055/s-0032-1312612. Epub 2012 Jul 19. Klin Padiatr. 2012. PMID: 22815129 Review.
Cited by
-
Accuracy of Pulse Oximetry in the Presence of Fetal Hemoglobin-A Systematic Review.Children (Basel). 2021 Apr 30;8(5):361. doi: 10.3390/children8050361. Children (Basel). 2021. PMID: 33946236 Free PMC article. Review.
-
Packed red blood cells transfusion in neonates: effect on FiO2 and PaO2/SaO2 ratio, and implications for neonatal saturation targeting.J Perinatol. 2018 Jun;38(6):693-695. doi: 10.1038/s41372-018-0079-x. Epub 2018 Feb 21. J Perinatol. 2018. PMID: 29467516
-
Lower versus higher oxygen concentrations titrated to target oxygen saturations during resuscitation of preterm infants at birth.Cochrane Database Syst Rev. 2018 May 4;5(5):CD010239. doi: 10.1002/14651858.CD010239.pub2. Cochrane Database Syst Rev. 2018. PMID: 29726010 Free PMC article.
-
Oxygen Supplementation During Preterm Stabilization and the Relevance of the First 5 min After Birth.Front Pediatr. 2020 Jan 31;8:12. doi: 10.3389/fped.2020.00012. eCollection 2020. Front Pediatr. 2020. PMID: 32083039 Free PMC article. Review.
-
Sickle Cell Disease and Fetal Hemoglobin.Saudi J Med Med Sci. 2018 Sep-Dec;6(3):131-132. doi: 10.4103/sjmms.sjmms_128_18. Epub 2018 Aug 14. Saudi J Med Med Sci. 2018. PMID: 30787838 Free PMC article. No abstract available.
References
-
- Wimberly PD. Oxygen monitoring in the newborn. Scand J Clin Lab Invest. 1993;54(Suppl 214):127–30. - PubMed
-
- Wimberly PD, Siggaard-Anderson O, Fogh-Anderson N. Accurate measurement of hemoglobin oxygen saturation, and fraction of carboxyhemoglobin and methemoglobin in fetal blood using Radiometer OSM3: Corrections for fetal hemoglobin fraction and pH. Scand J Clin Lab Invest. 1990;50(Suppl 203):235–9. - PubMed
-
- Shiao S-YPK. Functional versus fractional oxygen saturation readings: bias and agreement using simulated solutions and adult blood. Biosci Res Nur. 2002;3:210–21. - PubMed
-
- Shiao S-YPK. Effects of fetal hemoglobin on accurate measurements of oxygen saturation in neonates. J Perinat Neonatal Nurs. 2005;19:348–61. - PubMed
-
- Praud JP, Carofilis A, Bridey F, Lacaille F, Dehan M, Gaultier CL. Accuracy of 2 wavelength pulse oximetry in neonates and infants. Pediatr Pulmonol. 1989;6:180–2. - PubMed
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources