Monitoring of scoliosis surgery with epidurally recorded motor evoked potentials (D wave) revealed false results
- PMID: 16844406
- DOI: 10.1016/j.clinph.2006.05.021
Monitoring of scoliosis surgery with epidurally recorded motor evoked potentials (D wave) revealed false results
Abstract
Objective: To elucidate the mechanism behind D wave amplitude changes after surgical correction of scoliosis.
Methods: We collected D wave and muscle MEP data from 93 patients (78 female, 15 male, age range 4-19 years, mean age 15.9 years), who underwent surgical correction of scoliosis. D waves were recorded via a catheter electrode inserted epidurally through the flavectomy. Muscle MEPs from lower limb muscles were also recorded. Muscle MEPs/D wave were elicited by short trains/single transcranial electrical stimuli. SEPs were elicited through bilateral percutaneous stimulation of the tibial nerves at the ankle and an averaged response from 100 to 200 single sweeps were recorded over the scalp at Cz'/Fz. In addition, we analyzed intraoperatively obtained X-ray images of the spine in 9 patients and preoperative spinal MRI in two of those nine.
Results: After surgical correction of scoliosis in 25 of 93 (27%) patients, the D wave amplitude changed by more than 20% of its baseline value. A decremental change occurred in 21 (84%) and an incremental change in 4 (16%) patients. D wave decrements of more than 50% were observed in 5 patients without significant SEP changes in any of these cases. In 9 patients, intraoperatively obtained X-rays of the spine (before and after correction of spine curvature) showed no catheter displacement. Muscle MEPs did not change and postoperative sensory-motor status was normal. In 2 patients, preoperative MRI revealed displacement of the spinal cord towards the concave side of the scoliotic curvature.
Conclusions: During scoliosis surgery, D wave amplitude changes should be interpreted cautiously until the definitive cause(s) of these changes are found. One possible mechanism to explain D wave changes during scoliosis correction could involve rotation of the spinal cord within the spinal canal, and the relative position of the epidural recording catheter (ERC). Rotation of the spinal cord after correction of scoliosis could introduce a new relationship between the ERC and the corticospinal tracts (CTs). Due to high incidence of false D wave amplitude changes we suggest that this methodology should not be used to assess the functional integrity of the CTs during scoliosis surgery.
Significance: This study provides new insight into the methodology of D wave monitoring as well as strong evidence of a high incidence of false positive results using D wave monitoring during surgical correction of scoliosis.
Similar articles
-
The refractory period of fast conducting corticospinal tract axons in man and its implications for intraoperative monitoring of motor evoked potentials.Clin Neurophysiol. 2004 Aug;115(8):1931-41. doi: 10.1016/j.clinph.2004.03.016. Clin Neurophysiol. 2004. PMID: 15261873
-
Intraoperative neurophysiological monitoring of the spinal cord during spinal cord and spine surgery: a review focus on the corticospinal tracts.Clin Neurophysiol. 2008 Feb;119(2):248-64. doi: 10.1016/j.clinph.2007.09.135. Epub 2007 Nov 28. Clin Neurophysiol. 2008. PMID: 18053764 Review.
-
Spinal cord monitoring in scoliosis surgery using an epidural electrode. Results of a prospective, consecutive series of 191 cases.Spine (Phila Pa 1976). 2006 Oct 15;31(22):2614-23. doi: 10.1097/01.brs.0000240642.28495.99. Spine (Phila Pa 1976). 2006. PMID: 17047554
-
Multimodality monitoring of transcranial electric motor and somatosensory-evoked potentials during surgical correction of spinal deformity in patients with cerebral palsy and other neuromuscular disorders.Spine (Phila Pa 1976). 2003 Aug 15;28(16):1851-5; discussion 1855-6. doi: 10.1097/01.BRS.0000083202.62956.A8. Spine (Phila Pa 1976). 2003. PMID: 12923474
-
Intraoperative neurophysiologic monitoring for intramedullary spinal-cord tumor surgery.Neurophysiol Clin. 2007 Dec;37(6):407-14. doi: 10.1016/j.neucli.2007.10.003. Epub 2007 Nov 9. Neurophysiol Clin. 2007. PMID: 18083496 Review.
Cited by
-
A review of intraoperative monitoring for spinal surgery.Surg Neurol Int. 2012;3(Suppl 3):S174-87. doi: 10.4103/2152-7806.98579. Epub 2012 Jul 17. Surg Neurol Int. 2012. PMID: 22905324 Free PMC article.
-
Four-limb muscle motor evoked potential and optimized somatosensory evoked potential monitoring with decussation assessment: results in 206 thoracolumbar spine surgeries.Eur Spine J. 2007 Nov;16 Suppl 2(Suppl 2):S171-87. doi: 10.1007/s00586-007-0426-7. Epub 2007 Jul 19. Eur Spine J. 2007. PMID: 17638028 Free PMC article.
-
Intraoperative neurophysiological monitoring in spinal surgery.World J Clin Cases. 2015 Sep 16;3(9):765-73. doi: 10.12998/wjcc.v3.i9.765. World J Clin Cases. 2015. PMID: 26380823 Free PMC article. Review.
-
What Is the Best Multimodality Combination for Intraoperative Spinal Cord Monitoring of Motor Function? A Multicenter Study by the Monitoring Committee of the Japanese Society for Spine Surgery and Related Research.Global Spine J. 2016 May;6(3):234-41. doi: 10.1055/s-0035-1559582. Epub 2015 Jul 31. Global Spine J. 2016. PMID: 27099814 Free PMC article.
-
Current approach on spinal cord monitoring: the point of view of the neurologist, the anesthesiologist and the spine surgeon.Eur Spine J. 2007 Nov;16 Suppl 2(Suppl 2):S115-29. doi: 10.1007/s00586-007-0419-6. Epub 2007 Jul 10. Eur Spine J. 2007. PMID: 17619913 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials