Radiofrequency ablation (RFA): development of a flow model for bovine livers for extensive bench testing
- PMID: 16845558
- DOI: 10.1007/s00270-004-0338-4
Radiofrequency ablation (RFA): development of a flow model for bovine livers for extensive bench testing
Abstract
Purpose: To develop a flow model for bovine livers for extensive bench testing of technical improvements or procedure-related developments of radiofrequency ablation excluding animal experiments.
Methods: The perfusion of bovine livers directly from the slaughterhouse was simulated in a liver perfusion tank developed for the experimental work. The liver perfusion medium used was a Tyrode solution prepared in accordance with physiologic criteria (as for liver transplants) which was oxygenated by an oxygenator and heated to 36.5 degrees C. Portal vein circulation was regulated via a flow- and pressure-controlled pump and arterial circulation using a dialysis machine. Flow rate and pressure were adjusted as for the physiology of a human liver converted to bovine liver conditions. The fluid discharged from the liver was returned into the perfusion system through the vena cava. Extendable precision swivel arms with the radiofrequency probe attached were mounted on the liver perfusion tank. RFA was conducted with the RF3000 generator and a 2 cm LeVeen needle (Boston Scientific, Ratingen, Germany) in a three-dimensional grid for precise localization of the generated thermolesions.
Results: Four bovine livers weighing 8.4 +/- 0.4 kg each were prepared, connected to the perfusion system, and consecutively perfused for the experiments. Mean arterial flow was 569 +/- 43 ml/min, arterial pressure 120 mmHg, portovenous flow 1440 +/- 305 ml/min, and portal pressure 10 mmHg. Macroscopic evaluation after the experiments revealed no thrombi within the hepatic vessels. A total of 136 RF thermolesions were generated with an average number of 34 per liver. Mean RF duration was 2:59 +/- 2:01 min:sec with an average baseline impedance of 28.2 +/- 3.4 ohms. The mean diameter of the thermolesions along the puncture channel was 22.98 +/- 4.34 mm and perpendicular to the channel was 23.27 +/- 4.82 mm.
Conclusion: Extracorporeal perfusion of bovine livers with consecutive standardized RF ablation was feasible. The bovine liver flow model seems to allow extensive, standardized evaluation of technical or procedure-related developments of RF systems.
Similar articles
-
Effects of vascular perfusion on coagulation size in radiofrequency ablation of ex vivo perfused bovine livers.Invest Radiol. 2006 Apr;41(4):422-7. doi: 10.1097/01.rli.0000201231.60420.a2. Invest Radiol. 2006. PMID: 16523026
-
In vitro blood-perfused bovine liver model: a physiologic model for evaluation of the performance of radiofrequency ablation devices.J Vasc Interv Radiol. 2011 Oct;22(10):1478-83. doi: 10.1016/j.jvir.2011.06.013. Epub 2011 Jul 28. J Vasc Interv Radiol. 2011. PMID: 21802317
-
Effectiveness of hepatic arterial embolization on radiofrequency ablation volume in a swine model: relationship to portal venous flow and liver parenchymal pressure.J Vasc Interv Radiol. 2008 Nov;19(11):1646-51. doi: 10.1016/j.jvir.2008.08.015. J Vasc Interv Radiol. 2008. PMID: 18954767
-
The isolated perfused rat liver: standardization of a time-honoured model.Lab Anim. 2006 Jul;40(3):236-46. doi: 10.1258/002367706777611460. Lab Anim. 2006. PMID: 16803641 Review.
-
Radiofrequency ablation of bone with cooled probes and impedance control energy delivery in a pig model: MR imaging features.J Vasc Interv Radiol. 2007 Aug;18(8):1011-20. doi: 10.1016/j.jvir.2007.05.015. J Vasc Interv Radiol. 2007. PMID: 17675620 Review.
Cited by
-
Fast Estimation of the Vascular Cooling in RFA Based on Numerical Simulation.Open Biomed Eng J. 2010 Feb 4;4:16-26. doi: 10.2174/1874120701004020016. Open Biomed Eng J. 2010. PMID: 20448794 Free PMC article.
-
Electrochemical treatment: An investigation of dose-response relationships using an isolated liver perfusion model.Saudi J Gastroenterol. 2011 Sep-Oct;17(5):335-42. doi: 10.4103/1319-3767.84491. Saudi J Gastroenterol. 2011. PMID: 21912061 Free PMC article.
-
Evaluation of the in vivo efficiency and safety of hepatic radiofrequency ablation using a 15-G Octopus® in pig liver.Korean J Radiol. 2013 Mar-Apr;14(2):194-201. doi: 10.3348/kjr.2013.14.2.194. Epub 2013 Feb 22. Korean J Radiol. 2013. PMID: 23482995 Free PMC article.
-
Creation of an ex-vivo bovine kidney flow model for testing embolic agents: work in progress.CVIR Endovasc. 2021 Feb 3;4(1):20. doi: 10.1186/s42155-021-00210-0. CVIR Endovasc. 2021. PMID: 33534088 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources