Air displacement plethysmography can detect moderate changes in body composition
- PMID: 16855541
- DOI: 10.1038/sj.ejcn.1602482
Air displacement plethysmography can detect moderate changes in body composition
Abstract
Objectives: To determine the sensitivity of air displacement plethysmography (APD) for evaluation of changes in body composition in normal subjects.
Design: Comparison of measurements with and without oil or water loads.
Subjects and methods: Ten healthy volunteers were analyzed, without and with 1 l and 2 l of oil or water. The measured and true changes in fat mass and fat-free mass were compared by paired t-tests. A correlation study and a Bland & Altman procedure was performed on the 60 measurements of adiposity changes in 30 subjects carrying 0.5 l (n=8 x 2), 1 l (n=10 x 2) and 2 l (n=12 x 2) oil and water loads.
Results: Fat-free mass increased when the 10 subjects were carrying water. When they carried oil, fat mass increased, however, a approximately 0.5 kg increase of fat-free mass was also detected. Two liters loads led to distinct changes: +1.49+/-0.59 kg fat and +0.50+/-0.60 kg fat-free with oil and +0.37+/-0.57 kg fat and +1.70+/-0.56 kg fat-free with water (both P<0.001). Mixed loads (+1 l oil and 1 l water) led to detect +0.85+/-0.48 kg fat and +1.09+/-0.45 kg fat-free (both P<0.005 vs without load). For the 30 subjects analyzed thrice, measured changes in fat and fat-free mass were slightly underestimated (-15%, NS) but correlated with the true changes. Measured changes in adiposity were correlated with the true changes, with no bias as indicated by the Bland & Altman procedure.
Conclusion: APD detects approximately 2 kg changes in fat or fat-free mass in small populations.
Similar articles
-
Body composition of obese subjects by air displacement plethysmography: The influence of hydration.Obesity (Silver Spring). 2007 Jan;15(1):78-84. doi: 10.1038/oby.2007.533. Obesity (Silver Spring). 2007. PMID: 17228034
-
Estimates of percentage body fat in young adolescents: a comparison of dual-energy X-ray absorptiometry and air displacement plethysmography.Eur J Clin Nutr. 2003 Nov;57(11):1402-10. doi: 10.1038/sj.ejcn.1601702. Eur J Clin Nutr. 2003. PMID: 14576753
-
The use of air displacement plethysmography in children and adolescents with cystic fibrosis.Eur J Clin Nutr. 2004 Jul;58(7):985-9. doi: 10.1038/sj.ejcn.1601919. Eur J Clin Nutr. 2004. PMID: 15220939
-
Air-displacement plethysmography: here to stay.Curr Opin Clin Nutr Metab Care. 2005 Nov;8(6):624-9. doi: 10.1097/01.mco.0000171127.44525.07. Curr Opin Clin Nutr Metab Care. 2005. PMID: 16205463 Review.
-
Neonatal body composition by air displacement plethysmography in healthy term singletons: a systematic review.BMC Pediatr. 2019 Dec 12;19(1):489. doi: 10.1186/s12887-019-1867-y. BMC Pediatr. 2019. PMID: 31830946 Free PMC article.
Cited by
-
Telephone-delivered nutrition and exercise counselling after auto-SCT: a pilot, randomised controlled trial.Bone Marrow Transplant. 2014 Jun;49(6):786-92. doi: 10.1038/bmt.2014.52. Epub 2014 Apr 7. Bone Marrow Transplant. 2014. PMID: 24710562 Clinical Trial.
-
Insulin therapy and body weight, body composition and muscular strength in patients with type 2 diabetes mellitus.J Nutr Metab. 2010;2010:340570. doi: 10.1155/2010/340570. Epub 2009 Oct 21. J Nutr Metab. 2010. PMID: 20721344 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous