Time poor: rushing decreases the accuracy and reliability of blood pressure measurement technique in pregnancy
- PMID: 16867915
- DOI: 10.1080/10641950600745228
Time poor: rushing decreases the accuracy and reliability of blood pressure measurement technique in pregnancy
Abstract
Background: In pregnancy, absolute blood pressure (BP) limits define preeclampsia. Therefore, BP in pregnancy should be measured accurately and in accordance with accepted guidelines. Accuracy of BP readings determined by rate of cuff deflation was analyzed. This study also investigated the compliance of clinical staff at Royal Prince Alfred Hospital, Australia, to guidelines for BP measurement.
Methods: The study was an observational trial of 98 normotensive antenatal or recently postnatal patients. Two BP readings were taken, each with fast (>5 mm Hg/sec) and slow (<or=2 mm Hg/sec) descent of mercury and compared by Bland-Altman analysis. Also, BP techniques used by junior doctors, specialist obstetricians, and midwives were compared using a 9-point scale.
Findings: Australian national guidelines recommend slow descent of mercury. Fast descent underestimated the systolic BP by 9 mm Hg (95% confidence interval [CI], -23 to +5 mm Hg) (p < 0.001). Fast descent measured the diastolic BP within 2 mm Hg (95% CI, -10 to +14 mm Hg) (not different, p = 0.151). Accuracy of fast cuff deflation was 28% for systolic BP and 50% for diastolic BP for <5 mm Hg, and respectively, 64% and 68% for <10 mm Hg, 84% and 80% for <15 mm Hg and 91% and 87% for <20 mm Hg. Compliance with guidelines was greatest for specialists and midwives (p = 0.001) and their most commonly missed feature (76% to 100%) was slow cuff deflation.
Interpretation: Rapid cuff deflation underestimates the systolic BP compared to accepted guidelines (<or=2 mm Hg/sec). Medical and midwifery staff may not follow accepted guidelines for BP measurement, particularly with regard to rate of cuff deflation. Potential misdiagnosis and under-treatment of patients with hypertension may compromise pregnancy outcomes.
Similar articles
-
Validity and reliability of a novel slow cuff-deflation system for noninvasive blood pressure monitoring in patients with continuous-flow left ventricular assist device.Circ Heart Fail. 2013 Sep 1;6(5):1005-12. doi: 10.1161/CIRCHEARTFAILURE.112.000186. Epub 2013 Jun 27. Circ Heart Fail. 2013. PMID: 23811966
-
Accuracy of the Dinamap 1846 XT automated blood pressure monitor.J Hum Hypertens. 2002 Sep;16(9):647-52. doi: 10.1038/sj.jhh.1001463. J Hum Hypertens. 2002. PMID: 12214262
-
Rapid assessment of blood pressure in the obstetric day unit using Microlife MaM technology.Hypertens Pregnancy. 2007;26(1):31-7. doi: 10.1080/10641950601146558. Hypertens Pregnancy. 2007. PMID: 17454216
-
Measuring blood pressure using the mercury sphygmomanometer.Br J Nurs. 2005 Feb 10-23;14(3):145-50. doi: 10.12968/bjon.2005.14.3.17519. Br J Nurs. 2005. PMID: 15788933 Review.
-
The assessment of blood pressure in pregnant women: pitfalls and novel approaches.Am J Obstet Gynecol. 2022 Feb;226(2S):S804-S818. doi: 10.1016/j.ajog.2020.10.026. Epub 2021 Jan 26. Am J Obstet Gynecol. 2022. PMID: 33514455 Review.