Performance of the automated Abbott RealTime HIV-1 assay on a genetically diverse panel of specimens from London: comparison to VERSANT HIV-1 RNA 3.0, AMPLICOR HIV-1 MONITOR v1.5, and LCx HIV RNA Quantitative assays
- PMID: 16876263
- DOI: 10.1016/j.jviromet.2006.06.010
Performance of the automated Abbott RealTime HIV-1 assay on a genetically diverse panel of specimens from London: comparison to VERSANT HIV-1 RNA 3.0, AMPLICOR HIV-1 MONITOR v1.5, and LCx HIV RNA Quantitative assays
Abstract
Automated RNA extraction and quantitation of HIV-1 by real-time PCR offer potential advantages of efficient sample processing, improved sensitivity, expanded dynamic range and reduced contamination risk. In this study, plasma was collected from 100 HIV-1 infected patients visiting The Courtyard Clinic of St. George's Hospital in London, United Kingdom (UK). Viral loads measured using the automated Abbott RealTime HIV-1 assay (m2000sp sample preparation and m2000rt amplification and detection instruments) were compared to results obtained with Versant HIV-1 RNA 3.0 (bDNA), AMPLICOR HIV-1 MONITOR v1.5 (Monitor v1.5) and LCx HIV RNA Quantitative (LCx HIV) assays. Based on gag p24, pol integrase, and env gp41 sequences, the panel included 26 subtype A, 20 B, 27 C, 10 D, 1 CRF01_AE, 3 CRF02_AG and 13 recombinant viruses. RealTime HIV-1, bDNA, Monitor v1.5 and LCx HIV quantitated 82, 74, 82, and 83% of samples, respectively, with 82, 71, 69 and 80 of the 100 samples measured within the dynamic ranges. Viral loads were highly correlated with 99% of values within 1 log(10) copies/ml between tests. The automated m2000 system and RealTime HIV-1 assay can increase laboratory throughput, enhance overall efficiency and reduce operator-associated errors while providing reliable quantitation of genetically diverse strains of HIV-1.
Similar articles
-
Evaluation of performance across the dynamic range of the Abbott RealTime HIV-1 assay as compared to VERSANT HIV-1 RNA 3.0 and AMPLICOR HIV-1 MONITOR v1.5 using serial dilutions of 39 group M and O viruses.J Virol Methods. 2007 Apr;141(1):49-57. doi: 10.1016/j.jviromet.2006.11.026. Epub 2006 Dec 20. J Virol Methods. 2007. PMID: 17184853
-
Performance of the automated Abbott RealTime HIV-1 assay on a genetically diverse panel of specimens from Brazil.J Virol Methods. 2006 Jun;134(1-2):237-43. doi: 10.1016/j.jviromet.2006.01.012. Epub 2006 Feb 28. J Virol Methods. 2006. PMID: 16510195
-
Comparative evaluation of the Cobas Amplicor HIV-1 Monitor Ultrasensitive Test, the new Cobas AmpliPrep/Cobas Amplicor HIV-1 Monitor Ultrasensitive Test and the Versant HIV RNA 3.0 assays for quantitation of HIV-1 RNA in plasma samples.J Clin Virol. 2005 May;33(1):43-51. doi: 10.1016/j.jcv.2004.09.025. J Clin Virol. 2005. PMID: 15797364
-
Comparison of the Abbott Realtime HIV-1 and HCV viral load assays with commercial competitor assays.Expert Rev Mol Diagn. 2008 Jul;8(4):369-77. doi: 10.1586/14737159.8.4.369. Expert Rev Mol Diagn. 2008. PMID: 18598219 Review.
-
The significance of HIV viral load assay precision: a review of the package insert specifications of two commercial kits.J Int Assoc Physicians AIDS Care (Chic). 2002 Oct-Dec;1(4):134-40. doi: 10.1177/154510970200100405. J Int Assoc Physicians AIDS Care (Chic). 2002. PMID: 12942672 Review.
Cited by
-
Evaluation of the Cavidi ExaVir Load assay (version 3) for plasma human immunodeficiency virus type 1 load monitoring.J Clin Microbiol. 2009 Sep;47(9):3011-3. doi: 10.1128/JCM.00805-09. Epub 2009 Jul 15. J Clin Microbiol. 2009. PMID: 19605583 Free PMC article.
-
Evaluation of the Abbott investigational use only realtime HIV-1 assay and comparison to the Roche Amplicor HIV-1 monitor test, version 1.5.J Mol Diagn. 2009 Jul;11(4):347-54. doi: 10.2353/jmoldx.2009.080166. Epub 2009 May 21. J Mol Diagn. 2009. PMID: 19460935 Free PMC article.
-
Sequence conservation of the region targeted by the Abbott RealTime HBV viral load assay in clinical specimens.J Clin Microbiol. 2013 Apr;51(4):1260-2. doi: 10.1128/JCM.03003-12. Epub 2013 Jan 23. J Clin Microbiol. 2013. PMID: 23345287 Free PMC article.
-
Correlation between human immunodeficiency virus type 1 (HIV-1) RNA measurements obtained with dried blood spots and those obtained with plasma by use of Nuclisens EasyQ HIV-1 and Abbott RealTime HIV load tests.J Clin Microbiol. 2009 Apr;47(4):1031-6. doi: 10.1128/JCM.02099-08. Epub 2009 Feb 4. J Clin Microbiol. 2009. PMID: 19193847 Free PMC article.
-
Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays.J Clin Microbiol. 2009 Jul;47(7):2209-17. doi: 10.1128/JCM.01761-08. Epub 2009 May 6. J Clin Microbiol. 2009. PMID: 19420172 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical