A short plate compression screw with diagonal bolts--a biomechanical evaluation performed experimentally and by numerical computation
- PMID: 16893595
- DOI: 10.1016/j.clinbiomech.2006.06.001
A short plate compression screw with diagonal bolts--a biomechanical evaluation performed experimentally and by numerical computation
Abstract
Introduction: Decreasing the length of the side plate of the dynamic hip screw would theoretically allow a smaller surgical incision, a shorter surgical time, decreased operative blood loss and minimal periosteal stripping. A new design of a very short plate dynamic hip screw based on two diagonal screws has been developed. Our study compares the new design and the four-hole side plate in respect to mechanical properties and bio-mechanical outcomes utilizing the Finite Element Analysis method.
Methods: Four pairs of fresh frozen cadaveric femora were extracted from male corpses aged 25-43 years (mean 34.8). One femur of each pair was fixated by means of the new system and the other by means of the conventional design. Mechanical loading was applied to all four pairs. The decline which occurred during the periodical loadings and the breakage loads of fixated bones were measured. Mechanical performance and probability of failure was assessed by conducting a mathematical analysis using the finite element method.
Findings: The average deflection under excessive cyclic loading was 33% higher in the bones fixated with the very short plate-dynamic hip screw device than in those fixated with the conventional dynamic hip screw. The average load failure during the collapse-loading test was 3120N for the very short plate-dynamic hip screw as compared to 4160N for the regular device. Mechanical testing did not provide decisive results regarding failure. The mathematical analysis performed indicated that the maximal stress in the very short plate-dynamic hip screw reached values 3-4-fold higher than in the regular dynamic hip screw.
Interpretation: Although the new design offers a minimally invasive approach to subtrochanteric femur fracture fixation, it was found to have insufficient biomechanical performance resulting in high probability of mechanical failure. The authors believe that the finite element method may have the potential to serve as an additional clinical tool for performing surgical preplanning and assist in decision making.
Similar articles
-
Biomechanical comparison of two side plate fixation techniques in an unstable intertrochanteric osteotomy model: Sliding Hip Screw and Percutaneous Compression Plate.Clin Biomech (Bristol). 2007 Dec;22(10):1112-8. doi: 10.1016/j.clinbiomech.2007.07.016. Epub 2007 Sep 27. Clin Biomech (Bristol). 2007. PMID: 17900766
-
Minimally invasive screw plates for surgery of unstable intertrochanteric femoral fractures: a biomechanical comparative study.Clin Biomech (Bristol). 2008 Oct;23(8):1012-7. doi: 10.1016/j.clinbiomech.2008.04.018. Epub 2008 Jun 24. Clin Biomech (Bristol). 2008. PMID: 18579266
-
Biomechanical analysis of distal femur fracture fixation: fixed-angle screw-plate construct versus condylar blade plate.J Orthop Trauma. 2007 Jan;21(1):43-6. doi: 10.1097/BOT.0b013e31802bb372. J Orthop Trauma. 2007. PMID: 17211268
-
Alternative techniques in trochanteric hip fracture surgery. Clinical and biomechanical studies on the Medoff sliding plate and the Twin hook.Acta Orthop Scand Suppl. 2000 Oct;295:1-31. Acta Orthop Scand Suppl. 2000. PMID: 11116961 Review.
-
[Limited Contact Dynamic Compression Plate (LC-DCP)--biomechanical research as basis to new plate design].Orthopade. 1992 Feb;21(1):11-23. Orthopade. 1992. PMID: 1549332 Review. German.
Cited by
-
Dynamic Hip Screw Plate Length in Stable Intertrochanteric Fracture Neck of Femur: A Systematic Review.Cureus. 2022 Mar 14;14(3):e23138. doi: 10.7759/cureus.23138. eCollection 2022 Mar. Cureus. 2022. PMID: 35444892 Free PMC article. Review.
-
Design considerations for patient-specific bone fixation plates: a literature review.Med Biol Eng Comput. 2023 Dec;61(12):3233-3252. doi: 10.1007/s11517-023-02900-4. Epub 2023 Sep 11. Med Biol Eng Comput. 2023. PMID: 37691047 Review.
-
Different internal fixation methods for Hoffa-like fractures of the tibial plateau: a finite element analysis.Front Med (Lausanne). 2023 Jul 3;10:1172377. doi: 10.3389/fmed.2023.1172377. eCollection 2023. Front Med (Lausanne). 2023. PMID: 37465644 Free PMC article.
-
Strength analysis of clavicle fracture fixation devices and fixation techniques using finite element analysis with musculoskeletal force input.Med Biol Eng Comput. 2015 Aug;53(8):759-69. doi: 10.1007/s11517-015-1288-5. Epub 2015 Apr 8. Med Biol Eng Comput. 2015. PMID: 25850983
-
Trajectory of bolt and length of plate in femoral neck system determine the stability of femur neck fracture and risk of subsequent subtrochanteric fracture : a finite element analysis.BMC Musculoskelet Disord. 2023 Jun 7;24(1):465. doi: 10.1186/s12891-023-06579-4. BMC Musculoskelet Disord. 2023. PMID: 37280558 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical