Influence of relative humidity and ozone on the sampling of volatile organic compounds on carbotrap/carbosieve adsorbents

Affiliations

Influence of relative humidity and ozone on the sampling of volatile organic compounds on carbotrap/carbosieve adsorbents

Fabienne Palluau et al. Environ Monit Assess. 2007 Apr.

Abstract

By using a dynamic dilution system, the atmospheric measurement of 11 selected toxics VOCs (ethylene, acetylene, propene, 1-butene, 1,3-butadiene, 1-pentene, 1-hexene, benzene, toluene, ethylbenzene, m+p-xylene) from the list WHO of 1996 and TO-14 method of US EPA by preconcentration by thermal desorption (TD), analysis by gas chromatography (GC), identification and quantification with a flame ionisation detector (FID) was developed and validated in term of metrology, especially the techniques of sampling of these VOCs with adsorbents cartridges "Air Toxics" when used with an "UMEG sampler" equipped in the inlet with a nafion membrane. In particular the influence of climatic conditions (temperature and relative humidity) and the influence of chemical factors like ozone, on the representativity of sampling were studied. Experiments made with various humidities showed that the addition of a nafion membrane in the inlet of the sampling system was required. Without this membrane, losses of compounds were observed for RH >50%. With this membrane, storage for 2 weeks in a refrigerator, as for canisters, did not induce a loss of compounds. No significative decrease of concentrations of the studied VOCs after 14 days storage, which are known to react with ozone, were observed with an ozone concentrations of 55 ppb. One explanation is that nafion membrane, placed in the inlet of the sampler, will neutralize ozone before entering the sampling tubes. This observation is in accordance with literature which states that the sampling of VOCs on Carbotrap cartridges without ozone scrubber induce a loss of compounds.

PubMed Disclaimer

References

    1. J Chromatogr A. 2002 Jun 7;958(1-2):219-29 - PubMed
    1. Chemosphere. 2000 Jan;40(1):49-56 - PubMed
    1. J Chromatogr A. 1998 Jul 10;813(1):101-11 - PubMed
    1. J Chromatogr A. 2000 Aug 11;889(1-2):271-9 - PubMed
    1. J Chromatogr A. 1998 Jun 5;809(1-2):47-63 - PubMed

Publication types

LinkOut - more resources