Tracheal ulceration and obstruction associated with flexible Bivona tracheostomy tubes
- PMID: 16913349
- DOI: 10.1177/0310057X0603400410
Tracheal ulceration and obstruction associated with flexible Bivona tracheostomy tubes
Abstract
An adjustable length tracheostomy tube, Bivona, which assumes a curve on insertion, has a tendency to straighten itself in situ. The straightening force, measured electronically, was maximal (0.21 Newton) when a tube was bent 90 degrees. We observed particular clinical disadvantages of these tubes--that of tracheal ulceration (1 case), distortion of soft tracheal graft tissue (1 case) and airway obstruction when the tip embedded into the tracheal wall (1 case).
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical