Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 1999 Oct 15;7(4):e7.
doi: 10.3171/foc.1999.7.4.2.

Incidence and management of subdural hematoma/hygroma with variable- and fixed-pressure differential valves: a randomized, controlled study of programmable compared with conventional valves

Affiliations

Incidence and management of subdural hematoma/hygroma with variable- and fixed-pressure differential valves: a randomized, controlled study of programmable compared with conventional valves

P W Carmel et al. Neurosurg Focus. .

Abstract

Shunt systems with differential pressure valves are prone to the complications of overdrainage. A programmable valve permits adjustment of the opening pressure of the valve. In this paper the authors report the incidence of subdural fluid collections in a randomized trial of programmable compared with conventional valves, and they describe methodologies used in management of this complication. A multiinstitutional, prospective, randomized trial of the Codman Hakim programmable valve and conventional fixed-pressure valves was undertaken. Two classes were defined: "new" and "replacement" valves. Randomization of the type of valve in each group was performed at each study site. Clinical and radiological studies were required at fixed intervals over a 104-week period. All complications were reported. The experimental valves were required to be reprogrammed after magnetic resonance imaging studies, but all other decisions regarding pressure setting were left to each investigator. Three hundred seventy-seven patients were randomized; 194 were treated with a programmable valve and 183 with a fixed-pressure valve. The two groups were statistically similar in demographic composition, as were the "new" and "replacement" categories. The investigators made 540 valve pressure changes (five per patient; range one-41 changes). More than half of the reprogramming adjustments were made in the first 3 months postplacement; 70% were made within 6 months. More than half of all reprogramming adjustments were required in a group of 30 patients. Four treatment modalities were observed: 1) 30% of the fluid collections resolved spontaneously (25% in the patients with programmable valves and 36.3% in those with conventional valves) and were largely found to be hygromas in infants and children; 2) four subdural fluid collections were unresolved and under observation; 3) the subdural hematoma was drained and the shunt removed (in 8.3% of patients with the programmable valve and 36.3% of those with the control valve); 4) the pressure of programmable valve was raised in 58% of patients (seven of 12), and this increase in opening pressure was a feature used by investigators to affect treatment. There was no significant difference in the incidence of subdural fluid collections between the programmable and fixed-pressure valve treatment groups. The programmable feature provided a considerable advantage in treatment when subdural collections occurred.

PubMed Disclaimer