Beware the airway filter: deadspace effect in children under 2 years
- PMID: 16918654
- DOI: 10.1111/j.1460-9592.2006.01895.x
Beware the airway filter: deadspace effect in children under 2 years
Abstract
Background: Filters are increasingly used in breathing circuits as they protect the circuit from contamination and facilitate humidification of inspired gas. The use of filters, however, can augment the anatomical deadspace. This can be significant in children because they have much smaller tidal volumes.
Methods: Following institutional ethical approval, 20 healthy children <2 years of age who required tracheal intubation were recruited. Ventilation was adjusted to achieve an endtidal carbon dioxide (P(E)CO(2)) of 4.6 kPa (35 mmHg) when sampled at the tracheal tube (TT) adapter. Following a 10-min period of stabilization, an airway filter (22 ml) was introduced into the circuit. The respiratory rate (RR) was then adjusted to return P(E)CO(2) to 4.6 kPa (35 mmHg).
Results: A mean increase in ventilation of 1.42 (0.38) l x min(-1) was required to maintain a normal P(E)CO(2) level. Airway pressure and respiratory rate increased by 7.9 mmHg (4.6) and 19.8 breath x min(-1) (8.7) respectively. The P(E)CO(2) and partial pressure of inspired carbon-di-oxide (PiCO(2)) measured from the TT adapter were higher than measured from the filter port. The mean increase was 3.6 (1.6) mmHg for P(E)CO(2) and 5.9 (3.9) mmHg for PiCO(2).
Conclusion: Amplified deadspace from airway filters results in a significant increase in ventilation needed to maintain a normal P((E)CO(2) in children <2 years of age with normal lungs. Sampling of P((E)CO(2) and PiCO(2) from the filter significantly underestimates the effect of increased deadspace. The effect of increased deadspace may be predicted using a proposed mathematical model.
Similar articles
-
The effect of increased apparatus dead space and tidal volumes on carbon dioxide elimination and oxygen saturations in a low-flow anesthesia system.J Clin Anesth. 2008 May;20(3):170-4. doi: 10.1016/j.jclinane.2007.09.013. J Clin Anesth. 2008. PMID: 18502358 Clinical Trial.
-
The effects of breath-holds and Muller manoeuvres on upper airway carbon dioxide concentration in humans.Respiration. 2007;74(5):533-6. doi: 10.1159/000098884. Epub 2007 Jan 24. Respiration. 2007. PMID: 17259692
-
ProSeal as an alternative to endotracheal intubation in pediatric laparoscopy.Paediatr Anaesth. 2007 Apr;17(4):327-32. doi: 10.1111/j.1460-9592.2006.02127.x. Paediatr Anaesth. 2007. PMID: 17359400 Clinical Trial.
-
An evaluation of capnography monitoring during the apnoea test in brain-dead patients.Eur J Anaesthesiol. 2007 Oct;24(10):868-75. doi: 10.1017/S0265021507000725. Epub 2007 Jun 20. Eur J Anaesthesiol. 2007. PMID: 17579950 Clinical Trial.
-
Deadspace ventilation: a waste of breath!Intensive Care Med. 2011 May;37(5):735-46. doi: 10.1007/s00134-011-2194-4. Epub 2011 Mar 11. Intensive Care Med. 2011. PMID: 21394623 Review.
Cited by
-
The effect of a pediatric heat and moisture exchanger on dead space in healthy pediatric anesthesia.Korean J Anesthesiol. 2012 May;62(5):418-22. doi: 10.4097/kjae.2012.62.5.418. Epub 2012 May 24. Korean J Anesthesiol. 2012. PMID: 22679537 Free PMC article.
-
A comparison of the breathing apparatus deadspace associated with a supraglottic airway and endotracheal tube using volumetric capnography in young children.Korean J Anesthesiol. 2021 Jun;74(3):218-225. doi: 10.4097/kja.20518. Epub 2020 Nov 17. Korean J Anesthesiol. 2021. PMID: 33198431 Free PMC article.
-
An innovative pediatric Y-connector to minimize dead space, and maximize safety and surgical access.Saudi J Anaesth. 2012 Jan;6(1):77-8. doi: 10.4103/1658-354X.93063. Saudi J Anaesth. 2012. PMID: 22412786 Free PMC article. No abstract available.
-
Misinterpretation of carbon dioxide monitoring because of deadspace of heat and moisture exchanger with a filter in pediatric anesthesia: A case report.Medicine (Baltimore). 2018 Aug;97(35):e12158. doi: 10.1097/MD.0000000000012158. Medicine (Baltimore). 2018. PMID: 30170460 Free PMC article.
-
Evaluation the effect of breathing filters on end-tidal carbon dioxide during inferior abdominal surgery in infants and changes of tidal volume and respiratory rate needs for preventing of increasing end-tidal carbon dioxide.J Res Med Sci. 2016 Nov 7;21:115. doi: 10.4103/1735-1995.193506. eCollection 2016. J Res Med Sci. 2016. PMID: 28255323 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous